Fasten the excitation method changeover lever to a filter cube.
•
Insert a filter cube along the dovetail mount from the side opening
(Fig. 11) and push in as far as it will go.
•
Remove the changeover lever from the filter cube and install the two remaining
cubes in the same manner.
•
Remember the position of the filter cube so that you can insert the ID tag (supplied
with the filter cube) for the cube added.
•
Install frequently used filter cubes at the end of the dovetail mount in order to make
replacement easier.
•
If daiscopic (transmitted light) illumination is also to be used, install one diascopic
cube (DIA-ILL).
Secure filter cube clamp plate with Allen hex screws (Fig. 9)
Install and secure the filter cube cover on both the right and left sides with the knurled
retaining screws.