Panorama Exposures
18
Panorama Exposures
(1) Turn Main Switch On
Make sure that the i-Dixel application is running.
Press the top (
|
) of the main switch. The blue main LED will
light up to show that the unit is on.
Do not turn the main switch on if the patient is standing
near the unit. The arm will move and it could hit the
patient.
Emergency Stop Switch
In case of an emergency, press the Emergency Stop Switch.
This will stop the arm’s rotation, lift movement and X-ray
emission. Do not use this switch for anything. If the Emergency
Stop Switch is pressed, the lift will stop within 10 mm and the
arm rotation within 15 degrees.
After Pressing the Emergency Switch
Guide the patient away from the unit and turn the main switch
off. This will return the unit to a safe condition.
Turn the switch in the direction indicated by the arrow and
restart the computer. Then turn the main unit back on and check
that it is set for Panorama Exposure. If the unit cannot be
returned to a safe condition or will not operate, contact your
local dealer or J. MORITA OFFICE.
The image will be lost if the Emergency Stop Switch is
pressed during its transmission or if the main switch is
turned off.
Main LED
Summary of Contents for veraviewepocs 2d
Page 1: ...2022 02 21 Pub No X7C0 91001 503 EN Printed in Japan...
Page 4: ......
Page 107: ...103 Reference Axis Panoramic Cephalometric...
Page 116: ......