96
Vertical Height of Focal Spot
1,055 to 1,775 mm (Panoramic)
970 to 1,605 mm (Option)
1,125 to 1,775 mm (with Cephalometric, Upgradable Cephalometric)
1,040 to 1,605 mm (Option)
Patient Positioning
Auto focus using light sensor for distance measurement and electrically
operated positioning system
Patient Positioning Beam
Class 2 Laser. Accordance with 21CFR Part1040.10 and IEC60825-1.
Wavelength: typical 655 nm
Beam divergence: 120 mm ±10% in length, 0.8±0.2 mm in width @250 mm
Pulse duration and repetition rate: Continuous
Maximum energy output: 1 mW
Attenuation equivalent of Patient head stabilizer
Panoramic
less than 1.7 mmAl
Cephalometric
less than 1.7 mmAl
Leakage Technique Factors
80 kV, 600 mAs/h
(80 kV, 10 mA, duty cycle 1:59, for example 7.4 s exposure per 7 minutes 17 seconds cool-down period)
90 kV, 600 mAs/h
(90 kV, 10 mA, duty cycle 1:59, for example 4.9 s exposure per 4 minutes 49 seconds cool-down period)
Measurement Bases
Tube voltage: Actual X-ray radiation is monitored by Non Invasive Evaluator of Radiation Output.
Tube current: The mA is measured by monitoring current in the HT return line, which equals the tube current.
Exposure time: Exposure time is measured the open time of gate circuit which distributes high voltage to X-ray
tube head assembly, by counter (TP8 – TP GND on CPU1 PWB)
Collimator
Panoramic slit and Cephalo slit,
Imaging Device
Panoramic (without Cephalometric): Built-in
With Cephalometric :
Model CDCP700, PAN/CEPH cassette
Sensor:
CCD
Detail recognition:
0.096 mm pixel size for Superfine mode
0.144 mm pixel size for Fine high-Speed mode and Digital Cephalometric
mode
Image size:
Panoramic:
max. 3,000 × 1,536 pixel (Superfine mode)
max. 2,000 × 1,024 pixel (Fine high-Speed mode)
Cephalometric: max. 1,750 × 1,537 pixel