4)Air Turbine
a)Take the handpiece out of its holder and turn on the spray switch.
b)Take the handpiece off its main tube.
c)Hold the end of the tube over the basin and step on the foot pedal.
Run water through the tube for at least one minute.
d)Finally reconnect the handpiece and return it to its holder.
Repeat the above procedure for the second handpiece.
5)Micromotor
a)Take the motor out of its holder and turn its spray switch on.
b)Connect an attachment that has a spray function.
c)Hold the attachment over the basin and step on the foot pedal.
Run water through the motor for at least 2 minutes.
Outlet for water of the micromotor is on the side of
the connection cylinder. Let the water spout after
confirming the position of the outlet.
d)Finally return the motor to its holder.
Repeat the above procedure for the second micromotor*, if the unit
is equipped with one.
6)Ultrasonic Scaler
CHECK
The handpiece must always be attached before the scaler is vibrated.
>>Otherwise, excessive loads will be applied to the circuit board and other
components, and this could result in a breakdown or other malfunction.
a)Take the handpiece out of its holder and turn on the spray switch.
b)Hold the handpiece over the basin and step on the foot pedal. Con-
tinue to operate the instrument for at least 3 minutes.
c)When flushing is completed, return the handpiece to its holder.
Wipe moisture off enough for corrosion and pre-
vention after confirm the abnormality from O-ring, if
handpiece connector leakage.
CHECK
CAUTION
[6] Maintenance 5.