30
Splay ON(Micromotor)
Pedal Rotaion Splay
Step Variable ON
[5]Operation 6.
(3)Micromotor
1)Operation
Take the micromotor out of its holder and step on the foot pedal to
operate it.
When the spray is on, spray starts the moment the foot pedal is pressed.
3)Micromotor Overload Protection Circuit
Never press down with steady and excessive force or intentionally cause the
motor to stop by overloading it.
>>Repeated or frequent triggering of the micromotor’s overload protection circuit will
Cause it to overheat and result in damaging the motor.
A sudden and excessive increase in the load on the micromotor will
trigger the overload protection circuit, and the motor will temporarily
stop running.
This is to prevent overheating of the motor and the circuit board.
Release the foot pedal once and remove the cause of the load. Then
press the pedal again to restart the motor.
CAUTION
2)Rotation speed of variable range
The speed varies within each range of the right chart.
Select the speed range with the speed range switch on the dentist’s
switch panel.
See the right chart for the standard speed at each range.
During rotation, the air for cooling and blowing flows out (except UL
and L range)
Micromotor Speed
Estimates
Setting Range (min
-1
)
*1
UL 100 - 250 *2
(Ultra Low)
L 100 - 3,000 *2
(Low)
M 100 - 9,000
(Medium)
H 100 - 40,000
(High)
*1: Factory setting
*2: Tip Air Off(Factory setting)