Instructions for Use 2018-04-23
61
No.
Type
Explanation and Response
Reference
113
Memory back-up error
(EEPROM)
Memory for EEPROM has been erased.
Response: For proper laser output, the
characteristic values must be reset. Contact J.
MORITA OFFICE.
201
Pulse misses
Laser is skipping pulses
Response: The flash lamp is probably old and not
working right. Go to Menu and use Refresh
Lamp.
If this does not work, contact your local dealer or
J. MORITA OFFICE.
202
Communication error
Communication failure from panel to laser
control unit.
Response: Restart the unit.
If this does not work, contact your local dealer or
J. MORITA OFFICE.
204
Purge air error
Cooling air for the hollow waveguide not
detected.
Response: The hollow waveguide may be loose.
Tighten the connection ring and restart the unit.
If this does not work, contact your local dealer or
J. MORITA OFFICE. Cable could be damaged
if used as is.
205
Laser output too high
Laser output does not match set value.
Response: Restart the unit. If this does not
work, contact your local dealer or J. MORITA
OFFICE.
This happens if the
laser has not been
calibrated for some
time.
206
Sudden laser output
anomaly
Sudden deviation of laser output.
Response: Restart the unit. If this does not
work, contact your local dealer or J. MORITA
OFFICE.
208
Sudden output drop
Output suddenly dropped.
Response: Possible mirror damage.
Contact J. MORITA OFFICE.
Detected during start
up.
501
Time to replace
cooling water and
deionization filter
cartridge.
Time to replace the cooling water and the
deionization filter cartridge.
Response: contact your local dealer or J.
MORITA OFFICE. The AdvErL EVO could be
damaged if the both are not replaced on time.
Replace them as soon as possible, within 1 or 2
months.
Replace the cooling
water and the
deionization filter
cartridge once a year.
502
Flash Lamp is worn
out.
After 10 million shots the flash lamp should be
replaced as its performance will start to
deteriorate. After 20 million shots the lamp has
reached the end of its working life. Although it
can still be used errors will occur more and more
frequently; replace it right away. Go to the
Menu to check the total shot number for the lamp.
Go to the Menu and
check Lamp Shot
Number.