Instructions for Use 2018-10-22
20
(5)
Put the handpiece in its hanger.
♦
Be careful not to damage a contact tip when you put
the handpiece in its hanger.
♦
To avoid damaging the tip, place the handpiece so
that the tip faces the laser unit.
* Put the tips in the Tip Stand after taking them out of their
cases.
♦
The actual contact tip could be damaged if it is sticking up
when the top of the stand is closed.
•
The contact tip could come off if it is not screwed all the way on, and this could cause an injury if it
happened during treatment.
•
Contact tips wear out and must be replaced periodically. Inspect tips carefully before using them (see
below). Worn tips could overheat and injure the patient.
・
Do not use chipped or worn tips.
・
Do not use tips if the output seems lower than normal.
・
If the guide light is dim or does not appear at all, the tip may be damaged.
•
Tips are sharp and can cause injury; handle them with care.
•
Use only contact tips specified for use with the AdvErL EVO laser system.
•
When putting tips on and taking them off, turn the key off or put the unit in standby mode.
•
Always hold the knurled part of the tip to screw it on or off; never grip the metal pipe, which could
damage the tip..
•
Never emit a laser beam without having the handpiece and a contact tip installed.
•
Check the ends of tips and make sure they are free of blood and other contamination or debris.
Otherwise, they could overheat, especially if the tip air and spray are turned off. Overheated tips could
injure the patient.