The contact tip could come off if it is not screwed all the way on, and this could cause an injury if it
happened during treatment.
•
The worn-away contact tips must be replaced periodically. Inspect tips carefully before using them
(see below). Worn tips could overheat and injure the patient.
・
Do not use chipped or worn tips.
・
Do not use tips if the output seems to be lower than the normal.
・
If the guide light is dim or does not appear at all, the tip may be damaged.
•
Tips are sharp and can cause injury; handle them with care.
•
Use only contact tips specified for use with the AdvErL EVO laser system.
•
When putting tips on and taking them off, turn the key off or put the unit in standby mode.
•
Always hold the knurled part of the tip to screw it on or off; never grip the metal pipe, which could
damage the tip.
•
Never emit a laser beam without having the handpiece and a contact tip installed.
•
Check the ends of tips and make sure they are free of blood and other contamination or debris.
Otherwise, they might be overheated, especially if the tip air and spray are turned off. Overheated tips
could injure the patient.