background image

TDY

18 

Demand Moore Reliability  

Trimming the Input

Use this feature of the TDY to enhance the accuracy 

of the transmitter by precisely matching its actual 

reading of the input to its scaling of the either one 

or two input points.  Trim the output by following the 
directions below:

1.  Connect the transmitter to your PC using the 

setup shown in Figure 2.

2.  Click the radio button labeled “Input Trim On”.  
3.  Click “Start”, then “InpTrim”.

4.  Select whether you will trim two points (upper 

and lower) or just one point by clicking on the 

appropriate radio button in the “Trim Pnts” 

section.

5.  In the “Trim Lower” and “Trim Upper” boxes, 

type in the sensor’s value that you are 

trimming.

6.  Attach the sensor that you wish to trim to the 

transmitter and place the sensor in a calibration 

bath.  Adjust the bath until its temperature 

matches the “Trim Lower” value you have 

chosen.  Click “Trim Lower” and wait for it to 

capture the value.

7.  Attach the sensor that you wish to trim to the 

transmitter and place the sensor in a calibration 

bath.  Adjust the bath until its temperature 

matches the “Trim Upper” value you have 

chosen.  Click “Trim Upper” and wait for it to 

capture the value.

8.  Click on the button labelled “QuiInTrim”, then 

the “Stop” button, and finally the “Prog” button 

to send the trim values to the transmitter.  

Trimming the Output

The TDY has a function that allows precise control 

over its output.  Note that to activate this function, a 

unit must be connected in the setup to current meter, 

as shown in Figure 2.

To trim TDY output:

1.  Set all unit parameters on the TDY Configuration 

Screen as required  for your intended application.

 

When the Configuration File shown on the screen 

has been modified for your application, click “Stop” 

to stop measurement.  

2.  Click “OutTrim”.

3.  Select and move the “Zero Scale Output” or the 

“Full Scale Output” slide bar to adjust the TDY 

output. 

 

Monitor the output with a calibrated current meter 

or with a voltmeter measuring the drop across a 

precision, 250 Ohm resistor (see Figure 2).

4.  When both zero and full scale have been adjusted 

to satisfaction, press “QuitTrim”, then “Prog” to 

program the SIY with the new trim values.

Note:

 

Changing the settings shown on the screen 

in 

“Trim”

 mode DOES NOT CHANGE THE 

UNIT’S OPERATING PARAMETERS, it only 

updates the screen.  The settings are saved 

to non-volatile memory by using the “

Prog

” 

function.

 

The “TDY Status” area of the Configuration Screen 

will reflect the new output settings.

 

Figure 9.

 Trim Capture Box

Summary of Contents for TDY

Page 1: ...a n d M o o r e R e l i a b i l i t y All product names are registered trademarks of their respective companies User s Manual TDY PC Programmable Temperature Transmitter with Display No 235 740 00J M...

Page 2: ...Europe Australia 16650 Schoenborn Street North Hills California 91343 6196 U S A Tel 818 894 7111 Fax 818 891 2816 E mail info miinet com TOLL FREE 1 800 999 2900 www miinet com 1 Lloyds Court Manor...

Page 3: ...re 10 Gathering the Equipment Needed 10 11 PC Configuration Software Summary 12 Keeping What You ve Got Safe Using the Get Setup Function 13 Selecting Input Type Wiring etc 14 Scaling the Input also S...

Page 4: ...Connected TDY 23 Installing and Connecting the TDY 24 Connecting the TDY in an Application 24 Recommended Ground Wiring Practices 24 CE Conformity 24 Dimensions 25 26 Installation in Hazardous Locatio...

Page 5: ...by user designation inversely proportional to the input Available in your choice of explosion proof or NEMA 4X enclosures the TDY has certified protection in even the toughest environments Combine th...

Page 6: ...isplay is always two decimal places Range 99999 to 99999 Minimum Display Span 1 00 Operating and Storage Range Transmitter 40 C to 85 C 40 F to 185 F Display 20 C to 65 C 4 F to 149 F Relative Humidit...

Page 7: ...o 650 C 148 to 1202 F 200 to 630 C 328 to 1166 F 200 to 510 C 328 to 950 F 200 to 600 C 328 to 1112 F 200 to 630 C 328 to 1166 F 200 to 850 C 328 to 1562 F 80 to 320 C 112 to 608 F 50 to 250 C 58 to 4...

Page 8: ...explosion proof enclosure with two inch entry ports and a glass cover BH2TG 2 Hub explosion proof enclosure with two inch entry ports and a glass cover BH2MG 2 Hub explosion proof enclosure with two M...

Page 9: ...TDY Configuration Program on a PC Once the program is loaded and running nearly all of the operating parameters for the connected transmitter are shown on a single screen see Figure 3 This makes it ea...

Page 10: ...verything is setup as shown in Figure 2 and it is not possible to store a tag name in the memory of a transmitter without connecting that transmitter to the Configuration Program Table 3 lists the thi...

Page 11: ...Y COMM CABLE 803 040 26 FOR NON ISOLATED 803 039 26 FOR ISOLATED 804 030 26 FOR USB Thermocouple Millivolts Inputs 2 Wire RTD Direct Resistance Inputs 3 Wire RTD Direct Resistance Inputs 4 Wire RTD Di...

Page 12: ...Software and will be used often The names and functions of the individual command buttons may change when you press a button 5 Output These scroll bars control the scaling and damping values of the T...

Page 13: ...am Main Screen To save the Configuration File already resident in the transmitter s memory 1 Start the PC Configuration Program making sure that the TDY is connected as shown in Figure 2 2 Click on th...

Page 14: ...tton at the bottom center of the screen then on the Prog button This transfers the settings from the software to the memory of the connected transmitter CAUTION Any time a connected TDY is programmed...

Page 15: ...wnloaded Skip to Step 6 for instructions on getting the settings programmed into TDY nonvolatile memory 5 Select one of these options a Download the current settings into the connected transmitter s m...

Page 16: ...ay Full The display zero and display full show what values will actually be displayed If the pull down menu in Figure 7 and the radio button on Figure 8 both have Input selected then Display Zero will...

Page 17: ...Program Screen do not take effect in the unit until the Configuration File is downloaded Skip to Step 5 for instructions on getting the settings programmed into TDY nonvolatile memory 4 Choose from on...

Page 18: ...it for it to capture the value 8 Click on the button labelled QuiInTrim then the Stop button and finally the Prog button to send the trim values to the transmitter Trimming the Output The TDY has a fu...

Page 19: ...o this click on the Stop BW Stop Broken Wire Detection button when it is active on the bottom of the screen Setting Miscellaneous Functions Engineering Unit Readout Input Filtering Upscale or Downscal...

Page 20: ...Main Screen or enter a value directly into the space provided The default damping value is 0 seconds The maximum allowable damping value is 5 056 seconds Using the Loop Test The TDY Loop Test feature...

Page 21: ...l down menu to set Linearization ON Then click the CustTabl button in the lower left area of the Configuration Main Screen Figure 11 Linearization Must be set to ON to Create a Custom Table 3 Click on...

Page 22: ...ests filling in the EGU Engineering Units then the number of points and finally the values in the table 9 When the table is complete click on the QuitTabl button and answer Yes on the pop up dialog bo...

Page 23: ...e File button to bring up the File Management portion of the program 2 Use the Explorer like file display see Figure 13 to locate the file to be retrieved and click on the file to highlight it 3 Click...

Page 24: ...output signals from Moore Industries productsshouldbewiredusingashielded twisted pair wiring technique Shields should be connected to an earth or safety ground For the best shielding the shield should...

Page 25: ...Dimensions of the HP housing for the TDY unit 83 8 mm 3 30 in 118 1 mm 4 65 in 130 mm 5 12 in 111 8 mm 4 40 in 82 6 mm 3 25 in 63 5 mm 2 50 in C L Interior Diameter 81 3 mm 3 2 in 134 6 mm 5 30 in In...

Page 26: ...TDY unit PS PS 1 2 3 4 SDY 16 203 MA 65mm 2 55 in GND 1 2 NPT 102mm 4 00 in 84mm 3 31 in 68mm 2 69 in 129mm 5 06 in 10mm 38 in 53mm 2 09 in 124mm 4 90 in 66mm 2 60 in 25mm 1 00 in 102mm 4 00 in 119mm...

Page 27: ...lled as Division 2 equipment the TDY shall be mounted within a tool secured enclosure which meets the requirements of ANSI ISA 61010 1or C22 2 No 1010 1 and be capable of accepting the applicable wiri...

Page 28: ...all be installed in an enclosure which maintains an ingress protection rating of at least IP20 The communications port shall be programmed through Moore Industries Cable Part No 804 030 26 as describe...

Page 29: ...ale causes this message in the Output section of the Configuration Screen This condition is downloadable to the unit and causes transmitter output to rise proportionally in response to a drop on the i...

Page 30: ...selves on their ability to provide timely accurate and practical answers to your process instrumentation questions Factory phone numbers are listed on the back cover of this manual If problems involve...

Page 31: ...ng the intrinsically safe option It also in cludes guidelines for setting up zener barriers neces sary in these types of applications These diagrams must be used to augment the installa tion instructi...

Page 32: ...TDY 32 Demand Moore Reliability...

Page 33: ...TDY Demand Moore Reliability 33...

Page 34: ...y and not to Buyer s customers or the users of Buyer s products unless Buyer is an engineering contractor in which case the policy shall extend to Buyer s immediate customer only This policy shall not...

Reviews: