ISSUE C 26

th 

Sept 2008 

 

PCR650 

 

OPERATING AND MAINTENANCE MANUAL 

 
 
 
 
 
 
  
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

 
 
 

Summary of Contents for PCR650

Page 1: ...ISSUE C 26 th Sept 2008 PCR650 OPERATING AND MAINTENANCE MANUAL ...

Page 2: ...2 ...

Page 3: ...LARM 7 SYSTEM MONITOR 7 SECTION 4 8 FILTERS 8 PRE FILTER CHANGING 8 MAIN HEPA FILTER CHANGING 8 SECTION 5 9 MAINTENANCE 9 FUSES 9 LIGHTING 9 UV TUBE 9 AIRFLOW 9 LOW AIRFLOW ALARM 9 CIRCUIT DIAGRAM 10 SECTION 6 11 SERVICING 11 Warning This system must be used in compliance with these instructions and any repairs or maintenance carried out by qualified personnel For parts or service information plea...

Page 4: ...ean air Laminar Flow over the entire worksurface To further reduce the risk of cross contamination between samples the PCR cabinet is supplied with a time controlled UV light to sterilise the working area between samples 6 individual pre set time functions are provided as standard Circulaire Circulaire Circulaire Circulaire PCR PCR PCR PCR Ca Ca Ca Cabinet binet binet binet External Dimensions 650...

Page 5: ...e sited in a draught free position ASSEMBLY The cabinet is supplied fully assembled with the main HEPA filter and pre filter fitted TESTING COMMISSIONING A test certificate will be supplied for conformity to CE marking electrical test and airflow ...

Page 6: ...n of the cycle 1 Press 5 minutes 2 Presses 10 minutes 3 Presses 15 minutes 4 Presses 20 minutes 5 Presses 25 minutes 6 Presses 30 minutes The UV pre set timed cycle will start the fluorescent light will switch off and the UV lamp will switch on The red UV warning light on the membrane panel will flash to indicate the cycle selected 1 flash 5 minutes 2 flashes 10 minutes etc When the UV timing cycl...

Page 7: ...on again When an audio beep sounds release the buttons The counter will be reset and the system return to normal mode LOW AIRFLOW ALARM The airflow is continuously monitored while the cabinet is running An audible alarm will sound if the airflow drops below a preset level and the red low airflow warning light will flash on the control panel This is an indication that the Pre Filter requires replac...

Page 8: ...rm is activated Remove the thumbscrews on top of the cabinet and lift off the pre filter retainer Replacement pre filters may be obtained from Monmouth Scientific Part No PF 0023 MAIN HEPA FILTER CHANGING The HEPA filter should be changed by trained personnel Contact Monmouth Scientific for advice ...

Page 9: ...l panel and may be changed by removing the two securing screws from inside the cabinet and hinging the control panel open The UV light fuse 2A Type F is located on the control circuit board AIRFLOW WARNING This procedure requires adjustment of live circuits and should only be carried out by a qualified person The airflow should be set using a calibrated rotating vane anemometer The fan speed is ad...

Page 10: ...10 CIRCUIT DIAGRAM ...

Page 11: ...s Check replace pre filter Check airflow monitor and re calibrate if necessary Check safety interlocks and alarms Inspect electrical components lighting cables etc Issue test report and airflow certificate For parts or service information please contact Monmouth Scientific on 44 0 1278 458090 ...

Page 12: ...12 Monmouth Scientific Units 5 6 Kilnside East Quay Bridgwater Somerset TA6 4DB Email info monmouthscientific co uk Monmouth Scientific is a trading division of Monmouth Surgical Ltd ...