25
4.0 ADVANCED
SETUP
The ADVANCED SETUP menu may be password protected. This
menu is used to program the various operating parameters of the
recorder. This includes analog inputs, alarms, display settings, the
clock and various trigger modes.
All items on this menu have one or more sub-menus. To locate the
Advanced Setup menu, press the
MENU
button from any normal view
mode. The Advanced Setup menu will not be visible when the menu
first appears. Use the Up Arrow (
▲
) and Down Arrow (
▼
) buttons to
scroll and highlight
ADV. SETUP
and press the
MENU
button. This
will bring up the Advanced Setup menu shown on the left. Current
settings in sub menus are indicated by a
(check mark) along side
the item or by the item being highlighted.
4.1 CHANNEL
SETUP
This menu option enables the user to program the inputs of the recorder. Each channel can
be set up as an analog input – volts, currents, thermocouple or RTDs, or as digital inputs
from RPM or frequency measurements. There is no need to swap out modules since the
inputs are universal. Filtering may be applied to each channel for stability of readings or to
catch peak values if the display rate is slower than the sample rate. (The unit always samples
at 100 samples per second per channel. For increased accuracy each range may be
calibrated independently. Use the Up Arrow (
▲
) and Down Arrow (
▼
) buttons to scroll and
highlight
CHAN n SETUP
(n = A or B) and press the
MENU
button. The menu options are
shown expanded below.
INPUT TYPE
Note: Menu will differ depending on Input Type
selection
VOLTS
250mV
to 25.0Vdc
Select full scale input Voltage
Input is for DC Volts
mA
4-20 mA to
0 - 50mAdc
Select full scale input Current. Input is for DC
milliAmps. External Shunt of 250 Ohms is
required.
THERMOCOUPLE
J, K, T, E
Select required Thermocouple (T/C) type
RTDs
Pt385 or 392
Select required RTD Type
FREQUENCY
100 Hz to 10kHz
Select Frequency for Full Scale Range
SPEED
100 RPM to
600,000 RPM
Select Speed for Full Scale Range
OFF
Turns Channel OFF (toggles) =OFF
LO SCALE
Default value is low end of range.
For Offset. Use
▼▲►
keys to adjust
◄
key escapes,
MENU
key to Save and Exit
HI SCALE
Default Value is top end of range.
For Full Scale. Use
▼▲►
keys to adjust.
◄
key escapes,
MENU
key to Save and Exit
CHAN A SETUP..
CHAN B SETUP..
ALARMS..
RELAYS..
DISPLAY..
DATA CARD..
CLOCK..
BEEPER..
SAMPLE TRIG..
REC. TRIG..
EXT. INPUT..