Diagnosis with LED 

Potential cause of error 

What must be done? 

LED 

doesn’t

 blink green 

Wrong power supply 

Check the power supply

 12…30V DC

, the correct polarity and the cable 

connection 

Reset of MWF necessary 

Interrupt the power supply in order to restart the device  
 
If the LED is not blinking after the restart, please proceed with step 27.1   

LED blinks green 

DIP switch not in position 0 

Put 

all 

DIP switch levers in the upper position off/0 

Analysis device or display shows no or 
wrong values  

Measure current of the analogue output signal 

4..20mA

 with a multimeter 

1.

 If no current is measurable, interrupt the power supply in order to 

restart the device 

2.

  If no current is measurable after the restart, please proceed with step 

27.1   

3.

  If  the measured current correlates to the filling level (e.g. 12 mA at a 

filling height of  50 %), please check the analysis device, display or the 
connection 

LED blinks green and the 
measured current 

doesn’t

 

correlate to the filling level 

Probe touches silo wall, installed 
equipment or nozzle 

Remove probe from silo wall, installed equipment or nozzle and make 
sure that the distance to them is sufficient 

Cornice near the probe 

Remove cornice near the probe 

Bridges in the silo 

Remove bridges in the silo and avoid bridge building 

Larger caking to the probe 

Remove larger caking and clean the probe 

Probe buckled 

Straighten the probe or renew it 

Probe released from the device 

Install probe or renew it 

Tensioning weight  released from the 
rope 

Install tensioning weight or renew it 

Parameterisation not correct 

Adapt parameterisation after consultation of MOLLET 

27.1

 If the malfunction persists, the MWF has to be returned to MOLLET for review of the device function.

28. Returns

   

e. g. flammable,  toxic, caustic or    cancer-producing.

28.2

 Furthermore please state: 

   

- Chemical and physical characteristics of the the bulk goods or 

   

  liquid

28.1

 Remove all adherent material residues of  filling material from the 

   

MWF. Be aware of seal grooves and cracks where material residues 

   

could stick. 

   

In particular if the bulk goods or liquids has been dangerous to 

   

health,  

   

-   Operating time of the MWF

   

-  Description of the failure occurred 

   

- Description of the application

M   LLET

Füllstandtechnik

GmbH

10

     MWF-BA-10     03/20

  ©

 by MOLLET

27. Troubleshooting

Operating instruction           MOLLET   D-74706 Osterburken  Tel.  +49  6291 6440-0  Fax  +49  6291 9846

Microwave level measurement

MWF

Summary of Contents for MOLOSwave MWF2 Series

Page 1: ...measuring data 04 Maximum forces Mounting position 06 Installation advices 07 Connection plan Switching sequence Electrical connection Potential compensation 09 Installation advices Preparation of mounting Mounting 08 Troubleshooting Returns 10 Page M LLET Füllstandtechnik GmbH 03 20 by MOLLET MWF BA 01 01 MOLLET D 74706 Osterburken Tel 49 6291 6440 0 Fax 49 6291 9846 Operating instruction Operati...

Page 2: ...d are sealed 1 5 Consider the data of the name plate on the device 1 4 For the electrical connection take notice of the local and statutory rules and regulations and or the VDE 0100 Please read and obey these safety instructions and the complete operating manual Prior to the use of the device in potentially explosive atmospheres please read and obey the and observe the operating instruction Specia...

Page 3: ...sensor electronic the process connection with the feed through low lateral forces at the probe and liquids electronic through the feed through to the probe in the vessel with the filling good and returned The high frequency measuring signal will be transmitted by the sensor MOLLET D 74706 Osterburken Tel 49 6291 6440 0 Fax 49 6291 9846 Operating instruction 03 20 by MOLLET MWF BA 03 03 5 Applicati...

Page 4: ...e rope rod Measuring range analog M 4 mA lower current value uMG down 150 mm 10 mm Probe length L Reference point R to end of probe up 120 mm 120 mm 10 Technicalmeasuringdata max measuring range probe length 20 mA upper current value oMG Factory setting uMG 4 mA Top edge tensioning weight up to 3 0 m at 0 3 m of filling level 1 m s or max 0 03 of measured value or depending on customers request Me...

Page 5: ... Pressure in container p 1 bar 40 bar Bulk goods temperature Dielectric constant εr 1 8 below 1 8 on request Probe length Wire rope probe LS 1 0 m 20 0 m Rod probe LW 0 5 m 3 0 m SM2E G1 50 8 G2I 1 4301 SM3E G1 55 8 G3I 1 4301 Art No Thread For thread code SM1E G1 41 6 G1I 1 4301 G2I G1 50 Delivery including seals G3I G1 55 13 Process connection thread Thread code Thread G1I G1 46 14 Hexagonal nut...

Page 6: ... way that distance remains to plane metallic walls A 100 mm to concrete walls A 500 mm outside of plastic containers D 300 mm to metallic hoppers and bottoms E 150 mm Exception Probe will be fixed The probe must not touch metallic walls and bottoms By distances C D 300 mm a disturbance signal suppression has to be done see parameterisation manual Disturbance signal scan Mounting position should be...

Page 7: ...ottom edge of the floor Fixing of the wire rope probe could be necessary when wire rope probe touches sometimes the silo wall the cone internal installations or other metallic parts Smaller diameters and heights 200 mm could restrict the measuring capability wire rope probe gets closer than 500 mm to a concrete wall For fixing a threat M12 is provided at the bottom of the tensioning weight rope st...

Page 8: ...unted vertically into the silo or vessel 1 2 Screw the bush of the probe on the thread of the feed through MWF has to be mounted into empty silos or vessels in order to make sure that wire rope probe hang vertically and don t knot Use for handling and wearing of the MWF the hexagon or the lower section of the housing 18 Preparation of mounting In most cases the MWF is delivered completely preassem...

Page 9: ...urement device MWF can be recycled 26 Disposal Disposal of the MWF is subject to the environmental legislation of the respective country in effect for the operator s premises Check the terminal assignments for correctness and the cable glands for tightness Switch on the power supply Adaption of parameters only by changed process conditions and after consulting MOLLET recommended 24 Commissioning A...

Page 10: ...d make sure that the distance to them is sufficient Cornice near the probe Remove cornice near the probe Bridges in the silo Remove bridges in the silo and avoid bridge building Larger caking to the probe Remove larger caking and clean the probe Probe buckled Straighten the probe or renew it Probe released from the device Install probe or renew it Tensioning weight released from the rope Install t...

Page 11: ...e Dokumente TDR Sensor zur kontinuierlichen Füllstandmessung TDR sensor for continuous level measurement Typ Type MWF conforms with the following European directives D 74706 Osterburken Mikrowellen Füllstandanzeiger Microwave level indicator Qualitätssicherung Osterburken den 03 03 2020 Wolfgang Hageleit DIN EN IEC 60079 0 2019 DIN EN 60079 11 2012 DIN EN 60079 31 2014 Geschäftsführer managing dir...

Reviews: