VersaMax, SpectraMax 340PC384, 190, Plus 384 Microplate Readers User Guide
38
0112-0126 B
Item
Description
Endpoint
1 second single wavelength
Kinetic
2 seconds minimum interval between reads single wavelength
Measurement Time for Cuvettes (calibration off) SpectraMax Plus 384
Item
Description
Plate: normal scan
33*K nm/min (8-well strip)
21*K nm/min (16-well strip)
Plate: speed scan
135*K nm/min (8-well strip)
77*K nm/min (16-well strip)
Cuvette: normal scan
45*K nm/min
Cuvette: speed scan
130*K nm/min
Scan Speed (*K = wavelength interval) SpectraMax Plus 384
Item
Description
Read chamber
Isothermal when temperature regulation is not enabled
Range
4°C above ambient to 45°C when temperature regulation enabled.
The ambient temperature must be >20°C to achieve temperature
regulation at 45°C.
Resolution
± 0.1°C
Accuracy
± 1.0°C for plate chamber and cuvette chamber
Temperature uniformity
at equilibrium
± 0.5°C at 37°C
Chamber warm-up time
15–30 minutes (measured on air) after initiation of temperature
regulation
Temperature regulation
4 sensors
Drift
± 0.2°C (regulated)
Temperature regulation
diagnostics
Temperature regulation system is continuously monitored and updated
Evaporation
Plate lid required to minimize evaporative cooling
Recommended plate
Flat-bottom plates with isolated wells and lid
Temperature Regulation SpectraMax Plus 384