Chapter 4
4-8
SPECTRAmax
190 Microplate Spectrophotometer Operator’s Manual
Table 4.1. Error Codes, Messages, and Notes about the Errors
ERROR
CODE
ERROR MESSAGE
NOTES
100-199: Unrecognized Command Errors Sent from the Computer
100
command not found
Command string not recognized.
101
invalid argument
Command Argument not recognized.
102
too many arguments
Too many arguments after command.
103
not enough arguments
Missing arguments.
104
input line too long
Too many characters in the input line.
105
command invalid, sys-
tem busy
Instrument could not perform the give command because it was
busy doing another task. Example: Request a wavelength while
the monochromator is in motion.
106
command invalid, mea-
surement in progress
Instrument could not perform command because a measurement
was in progress
107
no data to transfer
Inputting
transfer
when there’s no data in the buffer
108
data buffer full
To many data sets in the buffer. Can be caused by setting up a
long kinetic and disconnecting computer, or if SOFTmax PRO is
preempted by another application.
109
error buffer overflow
More than 65 errors in the buffer,
clear
the buffer.
110
stray light
111
invalid read settings
200-299: Firmware Errors
200
assert failed
Firmware error.
201
bad error number
Firmware error.
202
receive queue overflow
Caused by external device sending too much data over serial port
and ignoring flow control.
203
serial port parity error
Parity bit error detected with incoming serial data.
204
serial port overrun error
Caused by host computer sending too much data and ignoring
the flow control signal.
205
serial port framing error
206
cmd generated too
much output
Firmware error.
Summary of Contents for SpectraMax 190
Page 8: ...viii SPECTRAmax 190 Microplate Spectrophotometer Operator s Manual...
Page 10: ...Chapter 1 1 2 SPECTRAmax 190 Microplate Spectrophotometer Operator s Manual...
Page 23: ...Instrument Description SPECTRAmax 190 Microplate Spectrophotometer Operator s Manual 1 15...
Page 24: ...Chapter 1 1 16 SPECTRAmax 190 Microplate Spectrophotometer Operator s Manual...
Page 26: ...Chapter 2 2 2 SPECTRAmax 190 Microplate Spectrophotometer Operator s Manual...
Page 30: ...Chapter 3 3 2 SPECTRAmax 190 Microplate Spectrophotometer Operator s Manual...
Page 34: ...Chapter 4 4 2 SPECTRAmax 190 Microplate Spectrophotometer Operator s Manual...
Page 44: ...Chapter 4 4 12 SPECTRAmax 190 Microplate Spectrophotometer Operator s Manual...
Page 45: ...Appendix A Printers and Cables Compatible Printers A 3 Cables A 3...
Page 46: ...Appendix A A 2 SPECTRAmax 190 Microplate Spectrophotometer Operator s Manual...
Page 48: ...Appendix A A 4 SPECTRAmax 190 Microplate Spectrophotometer Operator s Manual...
Page 50: ...Appendix B B 2 SPECTRAmax 190 Microplate Spectrophotometer Operator s Manual...
Page 52: ...Appendix B B 4 SPECTRAmax 190 Microplate Spectrophotometer Operator s Manual...