Maintenance and Troubleshooting
22
5000499 D
Moving the Scanner
If the 4100A Microarray Scanner must be moved within the lab, or
returned to the factory for service, the scanner must be prepared for
transport.
If you do not have the original shipping crate, contact us and a new
crate can be provided for an additional charge.
To move the scanner
CAUTION!
Always use the Park Scanner utility prior to moving the
scanner. Failure to park the scanner can result in misalignment or
damage of the optical system.
1.
Remove any slide from the slide holder.
2.
Close the GenePix Pro Software.
3.
Click
Start > Programs > Molecular Devices >
GenePix Pro > GenePix Utilities.
The GenePix Utilities window appears
4.
Click
Park Scanner for Shipping.
A message box appears informing you that the scanner has
been successfully parked.
CAUTION!
Always turn off power to the scanner using the power switch.
Do not turn off the power by unplugging the main power cable from the
power outlet.
5.
Turn off the power switch on the scanner.
6.
Unplug the main power cable from the back of the scanner and
the lab’s power outlet.
CAUTION!
Never unplug the USB cable unless the scanner has been
turned off and the main power cable disconnected.
7.
Unplug the USB cable from the back of the scanner and the
control computer.
CAUTION!
Do not tilt or stand the scanner on its end when moving it
within the lab, or transferring it from the bench to the shipping crate.
Tip!
Do not turn on the power switch after parking the scanner.