6 Troubleshooting Procedures
6.4 Recovering from Mechanical Problems in Flex Mode
The procedures in this section are intended for cases in which tips have jammed, or the
power fails while you are using the instrument in Flex mode, and you see an error
message on the computer screen indicating a problem with the instrument. You may
need to open the instrument in order to locate and correct the problem. The procedures
in this section include:
•
6.4.1: Assessing a Mechanical Problem
•
•
6.4.3: Evaluating the Tip Rack
•
6.4.4: Inspecting Inside the Fluidics Module
•
6.4.5: Removing the Pipettor Head
•
6.4.6: Expelling Undispensed Fluid from Tips
Warning!
Always make sure the power switch on the instrument is in the OFF position,
and remove the power cord from the back of the instrument, prior to removing any
interior panels.
Caution:
Do not touch or loosen any screws or parts other than those specifically
designated in these instructions. Doing so may cause misalignment and will void the
instrument warranty.
102
FlexStation II Operator’s Manual – Rev. D