6
To access the Analog Out connector, press down on the
compartment cover on the back of the EM500. Slide it
outward, away from the center of the instrument. Connects
to a voltmeter or a data acquisition system. It provides a DC
voltage proportional to the display reading. The output is
scaled for 2 Volts full scale for the range selected.
The Analog Output BNC's maximum output is 2.00 Volts.
Even though the EM500 LCD may exceed its selected full
scale under some conditions, the Analog Output value will never exceed 2.00 Volts.
Example:
The LCD reads 3.47 mJ on the 20 mJ range (about 17 percent of full scale). The analog output will read
0.347 VDC.
Located on the right side of the instrument, the Input BNC
connects to a Molectron J25 or J50 series pyroelectric
probe.