154
Reference Manual
IP2048/RM, Rev AA
Appendix F: Configuring Using HART
January 2015
F.4.8
Echo size (D911)
The size of the echo returned from a surface depends on the surface range, gas composition and
temperature, transmitter performance, in-tank conditions (turbulence, presence of surface
foam, and draughts), and other factors.
The received echo strength may vary from pulse-to-pulse but monitoring the read-only
parameter Echo Size (D911) indicates the latest echo strengths.
The indicated value is an averaged percentage of the last five echoes, with 100% representing a
saturated returned signal.
Field Communicator
To view the echo strength:
1.
From the
Home
screen, select
3: Service Tools
.
2.
Select
1: Alerts
.
3.
Select
2: Diagnostic Data
.
4.
Select
1: Target Data
.
5.
Select
2: Echo Size D911
.
6.
When finished, press
EXIT
to exit to the previous menu.
Mobrey MCU900 Series control unit
To view the echo strength:
1.
From the
Main Menu
screen, select
MONITOR
.
2.
Select the transmitter (e.g. “
Tx1:MSP400RH
”).
3.
Select
DIAGNOSTICS
4.
Select
Echo Size
.
5.
When finished, press the
ESC
key to exit to the previous menu.
Fast Keys
3, 1, 2, 1, 2
Echo Size D911
53 %
EXIT
HELP
(Field Communicator Screen)
Echo Size
D911
75%
(Mobrey MCU901 Screen)
Summary of Contents for MSP400RH
Page 2: ......
Page 4: ......
Page 16: ...6 Reference Manual IP2048 RM Rev AA Section 2 Transmitter Overview January 2015 ...
Page 54: ...44 Reference Manual IP2048 RM Rev AA Section 4 Starting up January 2015 ...
Page 80: ...70 Reference Manual IP2048 RM Rev AA Appendix A Reference Data January 2015 ...
Page 84: ...74 Reference Manual IP2048 RM Rev AA Appendix B Product Certifications January 2015 ...
Page 87: ...77 Reference Manual IP2048 RM Rev AA Appendix C Integrated Display Menus January 2015 ...
Page 90: ...80 Reference Manual IP2048 RM Rev AA Appendix C Integrated Display Menus January 2015 ...
Page 98: ...88 Reference Manual IP2048 RM Rev AA Appendix E Field Communicator January 2015 ...