Error codes
TROUBLE:
When a fault occurs, the LED of the mo-vis Omni Joystick will start to
flash. A long delay is followed by a number of flashes with a short delay
The error message depends on the connection type. Count the number
of flases and look up the according error message in the table below:
FLASH COUNT
REASON
REQUIRED ACTION
1
-
-
2
Connection cable/dri-
ver issues
Check/replace cable to wheel-
chair, and / or check / replace
sensor cable (if available), or
replace PCB
3
Power supply
Check cable and/or replace PCB
4
Joystick/sensor fault
Check/replace joystick
5
-
-
6
ADC - internal Analog
to Digital converter
Replace PCB
7
Test flag failed or Diag-
nostic failed
Redo tests and/or replace PCB
35
Multi Joystick Installation Manual 35