Instructions for use
16
CO2CELL Comfort 190_np_en 1301_mmm_V1.02
RO6
In line CO
2
reducing valve with gauge
(0.1 to 2 bar)
PNEU
Automatic change over unit for two CO
2
cylinders
TYGON
Tygon gas tubing, per meter
12
CLEANING
Electrical hazard. Do not spill water or
FOHDQLQJÀXLGVRYHUWKHLQQHURURXWHU
surfaces.
Disconnect before cleaning.
Completely dry the unit before
reconnecting the power.
12.1
OUTER
SURFACES
Always disconnect the incubator before cleaning.
It is recommended that the exterior of the incubator regularly
wiped clean. Use a non-abrasive soft cloth dampened with a
warm soapy water solution.
Make sure the exterior is thoroughly dry after cleaning.
12.2
INNER
SURFACES
It is advisable to clean the chamber regularly.
Use a 70 % isopropanol 30 % distilled water solution.
Apply using a sterile cloth so as not to introduce any
contamination.
Never use any of the following:
x
Any chlorine based detergents or bleaches.
x
Any acidic cleaning agents.
x
Sodium Azide
x
Iodine
x
Ferric Chloride
Wipe over the inner surfaces of the stainless chamber, all
the shelves including the underside of the shelves, the shelf
racks and the water tray.
The shelves, racking and water tray can be sterilised in an
autoclave if required.
Clean the glass door and the inner gasket seal.
Clean the inner stainless steel surfaces of the outer door
and the outer door gasket.
12.3
OXIDATION
AT
HIGH
TEMPERATURE
(
COMFORT
190
S
ONLY
)
When operated at 200 °C the action of atmospheric oxygen
may cause discolouration of the stainless steel surfaces.
This natural oxidation is not detrimental to the quality of the
incubator and is completely harmless.
13
SERVICE
13.1
WARRANTY
All MMM Comfort 190 incubators are supplied with a two
year manufacturer’s warranty from the date of installation.
7RUHJLVWHU\RXUZDUUDQW\SOHDVH¿OOLQWKH8VHU*XDUDQWHH
Registration Card and post to MMM Limited or to your local
distributor.
13.2
PREVENTATIVE
MAINTENANCE
This Culture Safe CO
2
Incubator is built to last for many
years, but to ensure that the unit continues to perform
to the highest level we recommend that you purchase a
preventative maintenance contract from MMM.
The incubator should be calibrated annually.
13.3
RETURNING
A
UNIT
Should your incubator require repair or maintenance then
FRQWDFW\RXUORFDOGLVWULEXWRULQWKH¿UVWLQVWDQFH
Returns to MMM will only be accepted with prior
authorisation from the Service Department.
We recommend that you keep all packaging.
When returning a unit please include the following
information:
x
Incubator model and serial number
x
Date of purchase
x
Authorisation number
x
Name of local distributor if not purchased from MMM
x
A detailed description of the fault (we can never have too
much information)
x
Where was the incubator located
x
Contact details of the person who reported the fault
x
$FRPSOHWHGGHFRQWDPLQDWLRQFHUWL¿FDWHSOXVDFRS\
faxed through in advance)
MMM cannot accept any returns without a completed
GHFRQWDPLQDWLRQFHUWL¿FDWH