MMM CO2 Series Instructions For Use Manual Download Page 15

This shows that the door locks have activated and that 
Operator 6 is logged on.

Press 

to 

proceed.

8.3 

STERILISE

 

CYCLE

The sterilise cycle has now started.  
A temperature graph is plotted to show 
the duration of the cycle. The y-axis 

automatically changes its scale to 205 °C

This shows that 
the heaters are 
operational

Abort cycle icon  

D

Section 8.5

 

The sterilise cycle consists of 3 distinct phases. They are: 

x

  Heat up.

During the heat up phase, maximum power is applied to 
the heating elements. CO

2

 control is disabled to protect the 

sensor and the door locks are activated.

x

  2 hour dwell.

The dwell phase starts once the chamber temperature 
reaches 200 °C.

Heater power is automatically reduced to regulate the 
temperature at 200 °C.

x

  Cool down.

After 2 hours at 200 °C power is removed from the heaters 
and the cooling phase begins. The chamber is allowed to 
cool naturally. When the chamber temperature reaches 
38 °C the door locks are deactivated and CO

2

 control is 

enabled.

The time taken for a full sterilisation cycle is 
dependent on the cool down phase.

As cooling relies on the heat loss through the chamber walls 
the time taken is affected by the location of the incubator 
and the ambient temperature.

So, an incubator placed under a worktop will take longer to 
cool than a freestanding incubator on a worktop.

We recommend that you carry out the sterilise cycle at the 
end of a working day. That way, when you return to work in 
the morning the incubator will be ready for you to use.

Instructions for use

  CO

2

CELL Comfort 190_np_en 1301_mmm_V1.02         

13

 

HIGH

 

TEMPERATURE

 

STERILISATION

Risk of damage to unit.
The Comfort 190 is not intended to be used as 
an autoclave and must not be used for sterilising 
any other equipment.

Only the shelves, racking and water tray can be 
left inside the incubator during sterilisation.

Risk of injury. Risk of burning. Internal surfaces 
reach 200 °C

The Comfort 190S incubator is equipped with a high 
temperature sterilisation function. This raises the chamber 
temperature to 200 °C for 2 hours.

Independent evaluation has shown this to be effective 
against all spores and bacteria commonly associated with 
laboratory incubator contamination.

8.1 

PREPARATION

Make sure that you have entered a user name and 
password as shown in 

D

Section 6.6.1

Empty and dry the water tray. This can be placed back 
inside the incubator.

Clear any samples or equipment from the shelves.

The shelves and racking can be left in place during the 
sterilisation cycle.

8.2 

STARTING

 

THE

 

CYCLE

Press and hold the sterilise button for 5 seconds:

7KH¿UVWVWHULOLVHF\FOHFKHFNOLVWRSHQV

Automatically carried 
out by controller.

When you have emptied the water tray, place it back inside 
the chamber and close the doors.

Press the grey tick to acknowledge this:

All checks are now 
complete.

Press 

 to 

proceed.

Summary of Contents for CO2 Series

Page 1: ...Instructions for use CO2 INCUBATOR CO2Cell 190 Comfort CO2Cell 190S Comfort with sterilization 200 C CO2 CELL Comfort 190_np_en 1301_mmm_V1 02...

Page 2: ......

Page 3: ...ENT CONDITIONS 5 4 2 SHELVES AND USABLE SPACE 6 4 3 WATER TRAY 6 4 4 GAS CONNECTION 6 4 5 ELECTRICAL CONNECTION 6 5 GETTING STARTED 7 5 1 START UP 7 5 2 SETTING TIME AND DATE 7 5 3 FACTORY PRESETS 7 5...

Page 4: ...VOLT FREE ALARM RELAY BMS 15 10 2 4 20 MA RETRANSMISSION RETX 15 10 3 RS 232 SERIAL COMMUNICATION 232 15 10 4 ETHERNET COMMUNICATION ETH 15 10 5 MULTI DOOR 8ID 15 11 ACCESSORIES 15 12 CLEANING 16 12 1...

Page 5: ...for working in a laboratory environment These safety instructions are intended to be used alongside any local safety guidelines rules and regulations If in GRXEW SOHDVH FRQVXOW WKH HDOWK DQG 6DIHW 2I...

Page 6: ...turbing the temperature or atmosphere of the chamber The incubator can be optionally equipped with a multi door These multi doors improve temperature and CO2 recovery DIWHU D GRRU RSHQLQJ 0XOWL GRRUV...

Page 7: ...ON AND AMBIENT CONDITIONS The incubator can be placed on a table or worktop 3OHDVH HQVXUH WKDW WKH VLWH LV VXI FLHQWO VWURQJ enough to carry the weight of the incubator MMM do not recommend placing th...

Page 8: ...bator Empty the water tray if the incubator is switched off otherwise condensation will occur 4 4 GAS CONNECTION Risk of injury Compressed gas Risk of damage to unit Observe relevant regulations for h...

Page 9: ...d perform a self check 5 2 SETTING TIME AND DATE After the controller has booted up the time and date screen will be shown Time and date are shown in the format hh mm ss and dd mm yy If the time and d...

Page 10: ...ature alarm Door open alarm High CO2 alarm SD card alarm Low gas pressure alarm In an alarm condition an audible buzzer sounds and the alarm icon changes to red Press this red alarm icon and the alarm...

Page 11: ...function DSection 7 On the main screen press the graph symbol In this instance we are viewing the temperature graph Press the y axis limits The values can be changed using the keypad Press Use to vie...

Page 12: ...re that displays the time and date of 100 historical events When 100 events have been logged the oldest event is over written Events include alarms alarm acknowledgment door open closed power on off S...

Page 13: ...the following data to the SD card Date Time Chamber temperature set point SP Chamber temperature process value PV Door set point SP Door process value PV Door set point SP Door process value PV CO2 se...

Page 14: ...card is removed then the screen will automatically change to this Instructions for use 12 CO2 CELL Comfort 190_np_en 1301_mmm_V1 02 7 2 CHANGING FILENAME RX FDQ HDVLO FKDQJH WKH OHQDPH WR VWDUW D QHZ...

Page 15: ...worktop We recommend that you carry out the sterilise cycle at the end of a working day That way when you return to work in the morning the incubator will be ready for you to use Instructions for use...

Page 16: ...0 C Press to continue to main screen 8 5 ABORTING THE STERILISE CYCLE Should you need to stop the sterilise cycle for any reason then press the abort icon Press The incubator now enters the cooling p...

Page 17: ...as 4 mA 0 C 20 mA 250 C The CO2 signal is scaled as 0 mA 0 CO2 20 mA 20 CO2 10 3 RS 232 SERIAL COMMUNICATION 232 56 FRPPXQLFDWLRQ LV WWHG DV VWDQGDUG 7KLV DOORZV IRU the SD card data to be remotely a...

Page 18: ...H TEMPERATURE COMFORT 190S ONLY When operated at 200 C the action of atmospheric oxygen may cause discolouration of the stainless steel surfaces This natural oxidation is not detrimental to the qualit...