MMM CO2 Series Instructions For Use Manual Download Page 1

Instructions for use

CO

INCUBATOR

CO2Cell 190 Comfort
CO2Cell 190S Comfort  

(with sterilization 200 °C)

                           

CO

2

CELL Comfort 190_np_en 1301_mmm_V1.02

Summary of Contents for CO2 Series

Page 1: ...Instructions for use CO2 INCUBATOR CO2Cell 190 Comfort CO2Cell 190S Comfort with sterilization 200 C CO2 CELL Comfort 190_np_en 1301_mmm_V1 02...

Page 2: ......

Page 3: ...ENT CONDITIONS 5 4 2 SHELVES AND USABLE SPACE 6 4 3 WATER TRAY 6 4 4 GAS CONNECTION 6 4 5 ELECTRICAL CONNECTION 6 5 GETTING STARTED 7 5 1 START UP 7 5 2 SETTING TIME AND DATE 7 5 3 FACTORY PRESETS 7 5...

Page 4: ...VOLT FREE ALARM RELAY BMS 15 10 2 4 20 MA RETRANSMISSION RETX 15 10 3 RS 232 SERIAL COMMUNICATION 232 15 10 4 ETHERNET COMMUNICATION ETH 15 10 5 MULTI DOOR 8ID 15 11 ACCESSORIES 15 12 CLEANING 16 12 1...

Page 5: ...for working in a laboratory environment These safety instructions are intended to be used alongside any local safety guidelines rules and regulations If in GRXEW SOHDVH FRQVXOW WKH HDOWK DQG 6DIHW 2I...

Page 6: ...turbing the temperature or atmosphere of the chamber The incubator can be optionally equipped with a multi door These multi doors improve temperature and CO2 recovery DIWHU D GRRU RSHQLQJ 0XOWL GRRUV...

Page 7: ...ON AND AMBIENT CONDITIONS The incubator can be placed on a table or worktop 3OHDVH HQVXUH WKDW WKH VLWH LV VXI FLHQWO VWURQJ enough to carry the weight of the incubator MMM do not recommend placing th...

Page 8: ...bator Empty the water tray if the incubator is switched off otherwise condensation will occur 4 4 GAS CONNECTION Risk of injury Compressed gas Risk of damage to unit Observe relevant regulations for h...

Page 9: ...d perform a self check 5 2 SETTING TIME AND DATE After the controller has booted up the time and date screen will be shown Time and date are shown in the format hh mm ss and dd mm yy If the time and d...

Page 10: ...ature alarm Door open alarm High CO2 alarm SD card alarm Low gas pressure alarm In an alarm condition an audible buzzer sounds and the alarm icon changes to red Press this red alarm icon and the alarm...

Page 11: ...function DSection 7 On the main screen press the graph symbol In this instance we are viewing the temperature graph Press the y axis limits The values can be changed using the keypad Press Use to vie...

Page 12: ...re that displays the time and date of 100 historical events When 100 events have been logged the oldest event is over written Events include alarms alarm acknowledgment door open closed power on off S...

Page 13: ...the following data to the SD card Date Time Chamber temperature set point SP Chamber temperature process value PV Door set point SP Door process value PV Door set point SP Door process value PV CO2 se...

Page 14: ...card is removed then the screen will automatically change to this Instructions for use 12 CO2 CELL Comfort 190_np_en 1301_mmm_V1 02 7 2 CHANGING FILENAME RX FDQ HDVLO FKDQJH WKH OHQDPH WR VWDUW D QHZ...

Page 15: ...worktop We recommend that you carry out the sterilise cycle at the end of a working day That way when you return to work in the morning the incubator will be ready for you to use Instructions for use...

Page 16: ...0 C Press to continue to main screen 8 5 ABORTING THE STERILISE CYCLE Should you need to stop the sterilise cycle for any reason then press the abort icon Press The incubator now enters the cooling p...

Page 17: ...as 4 mA 0 C 20 mA 250 C The CO2 signal is scaled as 0 mA 0 CO2 20 mA 20 CO2 10 3 RS 232 SERIAL COMMUNICATION 232 56 FRPPXQLFDWLRQ LV WWHG DV VWDQGDUG 7KLV DOORZV IRU the SD card data to be remotely a...

Page 18: ...H TEMPERATURE COMFORT 190S ONLY When operated at 200 C the action of atmospheric oxygen may cause discolouration of the stainless steel surfaces This natural oxidation is not detrimental to the qualit...

Page 19: ...midity control type Passive Range rH 90 5 Door recovery 30 s opening minutes 50 ELECTRICAL DATA Nominal voltage 50 60 Hz V 230 1N PE Power Comfort 190 W 380 Power Comfort 190S sterilise cycle W 1800 P...

Page 20: ...Instructions for use 18 CO2CELL Comfort 190_np_en 1301_mmm_V1 02...

Page 21: ......

Page 22: ......

Page 23: ......

Page 24: ...0 1 2 3 4 5 6 4 7 8 9 1 7 2 8 1 3 8 4 0 2 3...

Reviews: