factor, i.e. the ratio of peak and RMS. If the crest factor exceeds 9 the additional
evaluation method should be applied.
If the worker being tested is sitting in the right position and vibration exposure has
started measurement can begin.
Select crest factor mode by pressing the ◄ key and start crest factor measurement
by the ► key (Reset). This will 

reset the RMS values of X/Y/Z, the vector sum a

W

 and MTVV to zero

restart the measurement timer.

Always press Reset before a measurement

  to establish the defined start condi-

tions.
If the displayed crest factors X/Y/Z are below 9, switch to Interval RMS mode by
pressing the ◄ key. At crest factors above 9 you should measure in addition vibra-
tion dose as shown in the next section.
Interval RMS values and crest factors are calculated based on the elapsed measuring
time since the last pressing of the key ► (Reset). This means that the fluctuation of
the displayed values becomes lower the longer the measurement lasts. Single peaks
will have less influence than at the beginning. Relevant for evaluation is only the
value at the end of measurement. The recommended measuring time for whole-body
vibration is at least 2 minutes. To alert you, the timer in the upper right corner re-
mains red until 2 minutes have elapsed.
The displayed interval RMS for X/Y/Z and vibration total values a

w

  include the

weighting factors (k) according to table  2. The weighting factors are set to 1 for
maximum RMS (MTVV) and vibration dose value (VDV).
“OVERLOAD” is indicated instead of the measuring values if the current magni-
tude is too high. Even if the overload condition was of short duration the measured
interval RMS value may become invalid because of missing samples. An overload
event during the entire averaging time is indicated by “OVL!” in the upper right
corner after the date. This warning can be deleted by pressing the key ► (Reset).
After measurement you may save the results by pressing the key ▼. Measurement
should be either continued or finished immediately before pressing the key. Other-
wise the measuring values will drop slowly. You will be asked to enter two lines of
ten capital letters or numbers as a comment (Figure  12, page  11). Use the keys
◄▲▼►  to select characters and to change the input position. Press F1 to change
the input line. Measurement can be finished before entering a comment because the
results will have already been saved by pressing the key ▼. 
Measurements can only be saved if a sensor is detected at X/Y/Z and if there is no
overload condition. In these cases the VM31 will display “Sensor error” or “Over-
load occurred” instead of saving in order to avoid invalid recordings. If an overload
occurred since the last reset (“OVL!” in upper right  corner) the instrument  will
show a warning “Overload occurred after last reset! Save anyway?”. 
If you want to measure several partial exposures you may now do further measure-
ments (see section 3.2).
For the calculation of 

vibration exposure A(8

) and to store results on a PC the Ex-

cel macro file vm31.xlsm is provided (see section 13.3).

17

Summary of Contents for VM31

Page 1: ...ed Weber Metra Mess und Frequenztechnik in Radebeul e K Meissner Str 58 D 01445 Radebeul Tel 49 351 836 2191 Fax 49 351 836 2940 Email Info MMF de Internet www MMF de Human Vibration Analyzer VM31 Val...

Page 2: ...Fax 49 351 836 2940 Email Info MMF de Internet www MMF de Note The latest version of this manual can be found at http www mmf de prod uct_literature htm Specification subject to change 2014 Manfred W...

Page 3: ...with VDV Values 18 4 4 2 3 Seat Effective Amplitude Transmissibility SEAT 18 5 General Vibration Measurement 19 6 Data Logger 21 7 Frequency Analysis 22 8 Data Memory 23 8 1 Measurement Data Memory 23...

Page 4: ...2631 and the EU Directive 2002 44 EC A fourth measuring channel can be used for example for SEAT mea surements seat effective acceleration transmissibility The VM31 meets the requirements for human vi...

Page 5: ...on of human exposure to hand transmitted vibration ISO 2631 Evaluation of human exposure to whole body vibration also ASA ANSI S3 18 ISO 8041 Human response to vibration Measuring Instrumentation ISO...

Page 6: ...oducing the least possible vibration Provision of auxiliary equipment that reduces the risk of injuries such as protec tive gloves or special seats Appropriate maintenance programs for work equipment...

Page 7: ...ral partial exposures with different vibration magnitudes This can be the case if there are longer interruptions in the work process if the work equipment or its way of use is changed A partial exposu...

Page 8: ...e are based on RMS values An alternative method uses fourth power mean values called Vibration Dose Value VDV with the mea suring unit m s1 75 Table 1 also includes VDV based limit values VDV based da...

Page 9: ...moving the batteries To insert the batteries remove the two screws from the back cover of the device and open the battery compart ment Figure 2 When inserting the batteries please en sure that their p...

Page 10: ...ble TEDS version the VM31 will open a menu for entering the sensitivities manually see section 10 1 The sensitivities entered remain stored as long as the sensor is connected even if the batteries are...

Page 11: ...performed with the same hand pressure force as used under normal operating conditions Since most machine tool handles do not provide surfaces for the adhesive or screw attachment of sensors Metra off...

Page 12: ...RMS Maximum Transient Vi bration Value MTVV which may indicate the pres ence of shock vibration The frequency weighting for hand arm vibration is Wh Figure 11 shows the filter of the VM31 and the tol...

Page 13: ...sing samples An overload event during the entire averaging time is indicated by OVL in the upper right corner after the date This warning can be deleted by pressing the key Reset After measurement you...

Page 14: ...lt into a flat rubber pad which adapt themselves to the interface between the vibration source and the test person Figure 4 The following measuring points are suitable On the seat surface under a seat...

Page 15: ...ll always be in a vertical po sition with Z perpendicularly to the backbone However to compensate this the VM31 auto matically swaps the X and Z axis for backrest measurement 4 4 2 VM31 Settings 4 4 2...

Page 16: ...other postures sensor positions and differ ent frequency weightings but the general procedure is the same Figures 17 to 20 show the frequency response curves of the weighting filters for comfort measu...

Page 17: ...ger trains 0 1 1 10 100 1000 0 0001 0 001 0 01 0 1 1 10 Wb min ISO 8041 Wb max ISO 8041 Wb VM31 Hz Figure 18 Whole body weighting filter Wc for the backrest of seats 0 1 1 10 100 1000 0 0001 0 001 0 0...

Page 18: ...evaluation can be the running RMS method MTVV or the fourth power vibration dose method VDV whereupon the latter is more common One of the two additional evaluation methods should be applied for occas...

Page 19: ...et to 1 for maximum RMS MTVV and vibration dose value VDV OVERLOAD is indicated instead of the measuring values if the current magni tude is too high Even if the overload condition was of short durati...

Page 20: ...g the key Reset Measurements can be saved by pressing the key see page 17 For vibration exposure VDV 8 calculation the duration of VDV measurement should be noted see section 3 2 The Excel file vm31 x...

Page 21: ...ssing the key Please note that the RMS and peak values in the general vibration ranges are mea sured over the last display cycle not over a long period A display cycle can be between 1 and 4 seconds d...

Page 22: ...lacement Low frequencies including noise are strongly amplified High pass filtering is inevitable The frequency response curves of the general vibration ranges can be seen in Figures 27 28 and 29 20 F...

Page 23: ...he Excel file vm31 xlsm can be used to transfer the measurements to a PC 6 Data Logger The VM31 can automatically record measuring data for channels X Y Z at the speed of the display re fresh rate It...

Page 24: ...r will be halted while data is transferred via the USB interface This may result in missing records The automatic shut off function see chapter 10 3 is not in use while logging The maximum number of r...

Page 25: ...nt data can be transferred to a PC using the Excel macro file vm31 xlsm 8 2 Logger Data Memory To view stored logger data press F3 to open the menu and select Data memory Data logger In this menu open...

Page 26: ...check or to change the entered sensitivities Sensitivities are entered one after another for chan nels X Y Z and if connected for channel A The sensitivity value is a five digit number with the measu...

Page 27: ...ypes To select the battery open the main menu by pressing F3 select Device settings and Battery type Within the sub menu select Battery type Figure 40 pressing selects between Alkaline non rechargeabl...

Page 28: ...PC The VM31 has a USB interface The VM2x USB cable Figure 44 is connected to the VM31 via its 8 pin connector Switch the device off before connecting it to the PC Connect the other end of the cable t...

Page 29: ...8 can be calculated from the human vibration records and measuring reports can be generated FFT data can also be transferred and displayed graphically To work with the file you need to allow macro ex...

Page 30: ...tion Exposure A 8 and VDV 8 Vibration exposure A 8 and alternatively VDV 8 are both used for the risk as sessment of human vibration They can be calculated based on hand arm H A and whole body measure...

Page 31: ...ctivity Below you will find the vi bration exposure results for each person Figure 48 13 4 FFT Data Import to Excel The FFT data stored on the VM31 can also be transferred to the Excel macro file Swit...

Page 32: ...on the start screen Figure 50 If a Firmware version with a higher number is available on the website proceed as follows 1 Download the firmware file vm31 hex from the above named internet address 2 Al...

Page 33: ...Figure 52 Figure 52 Firmware update 7 Click on Send in the Firmware Updater Transfer of the Firmware data has now begun The transfer progress is displayed as a time bar on the PC and also on the VM31...

Page 34: ...erator needs to be connected via the shown R C combination to the inputs X Y Z and A The 4 7 k resistor serves as a sink for the IEPE constant cur rents The 1000 F capacitor keeps the resulting DC vol...

Page 35: ...mV ms 2 0 01 m s 0 1 mm s 1 m Sensor with 100 mV ms 2 0 001 m s 0 001 mm s 0 1 m Linearity range 75 dB for 6 error Noise 0 003 m s Filters Human vibration Acceleration Velocity Displacement Weighting...

Page 36: ...er this warranty does not apply to any improper or inadequate maintenance or modification and operation outside the product s specifications Shipment to Metra will be paid by the customer The repaired...

Reviews: