General Specifications
Water
The Tsunami is a closed loop water cooled system and requires no utility water services. Do not
place the chiller above the laser. Should the unit not be installed properly, and a leak develops,
dripping w a t e r may damage one or both lasers. The hoses are 3.6 M (12ft) long. Connect the
Chiller output first to the Millennia, from the Millennia to the Tsunami then back to the chiller.
You may have the newer model Millennia eV that comes with cooling hoses that splits in parallel to
cool both components.
Power Requirements
Electrical
3955/3930
1
3955/3930
1
Chiller
Type
Single Phase
2
Single phase
2
Single phase
2
Voltage
110 VAC
220 VAC
110 / 220 VAC
Current
<1A
<0.5A
10A
Footnote:
1 – With optional Lok-To-Clock configuration.
2 – With earth ground.
Power connections
Before connecting the power supply to the AC power line, make sure that all the power supply
power switches are
OFF.
Final Check:
Upon the completion of the installation of utilities, verify that the services meet
all building safety codes.
SPECTRA-PHYSICS’ SERVICE ENGINEERS CANNOT PERFORM
ELECTRICAL OR PLUMBING WORK AT YOUR SITE.