Maintenance Agreement 

To maintain a valid warranty on your Spectra-Physics system, it is necessary for the 
customer to assume the responsibility and perform the routine maintenance program. 
Failure to do so may result in the warranty being voided. 

 

Nalco Mixture 

 

The Nalco Cleaner (P/N 1607-0547) is used to flush your system clean of ferrous metals 
and copper alloys as a result of corrosion. It is recommended as a maintenance 
procedure that the chiller/critical cooling loop in each piece of equipment is cleaned and 
treated 1x per year. The laser does not have to be lasing. The cleaner should be 
circulated for a minimum of 8 hours (the longer the better if time permits in order to 
assure thorough cleaning). 

 
 

Filters 

 

 

ALPS filter 

 

All models except HighQ-2 and Spirit One: 

When the laser cavity humidity reaches an unacceptable level, the ALPS air purge filter 
(P/N 90074765) should be replaced. This filter has an average lifetime 1.5 to two years, 
depending on the relative  humidity and the temperature of the operating environment. 

 

All models except HighQ-2: 

The Chiller’s filter screen should be inspected and cleaned every 3 months, or else 
exchanged when it can no longer be rinsed clean. You should also monitor the laser 
Chiller temperature to make sure cooling is within specifications. 

If the laser is not going to be used for an extended period of time, turn off the laser and 
drain the cooling solution. 
 

Summary of Contents for Spectra-Physics 100

Page 1: ...Newport Corporation MKS Spectra Physics Confidential 1 of 11 Pre Installation Guide For all Spirit HighQ 2 Laser Models Spectra Physics Inc 3635 Peterson Way Santa Clara CA 95054 April 2020...

Page 2: ...Instruction Manuals 4 Laser Safety Considerations 4 Diagnostics 4 Pre Installation Considerations 5 Location and Environment 5 Physical Description 6 Utility Requirements 7 Computer Requirements 8 Pr...

Page 3: ...d to carefully consider your operating environment prior to installation Proper water power and room temperature are required for each system A checklist of pre installation considerations is provided...

Page 4: ...ction Manuals Please read the manual to get vital information about your system Familiarize yourself with the system You are encouraged to spend as much time as possible reviewing the system component...

Page 5: ...ons ambient room temperature changes may be an important factor for the laser system s performance Air ducts should not blow directly on laser or optical path E To ensure stable day to day operation t...

Page 6: ...x 6 4 683 x 310 x 163 88 0 40 0 1040 8 SHG Laser Head 29 9 x 12 2 x 6 4 759 x 310 x 163 99 0 45 0 Power Supply 11 0 x 5 7 x 1 9 280 x 144 x 48 5 8 60 3 9 Chiller 15 7 x 19 0 x 10 5 400 x 483 x 267 68...

Page 7: ...ments HighQ 2 Controller 100 240 VAC 50 60 Hz Current 2A Spirit One Power Supply 90 305VAC 47 63 Hz Current 10A Chiller HECR008 100 240VAC 50 60 Hz Current 10A Spirit 4 Spirit 8 Spirit 16 non HE Contr...

Page 8: ...utilize the laser s built in GUI via direct connection to the Spirit head s ethernet port RJ 45 connector HighQ 2 Spirit One Spirit 4 Spirit 8 Spirit 16 Spirit 16HE Spirit 30HE Spirit graphical user i...

Page 9: ...e the system on your work surface Two or more people may be required to lift some equipment Compare the packing list with your quotation Call your Spectra Physics office about any discrepancies Check...

Page 10: ...ing The cleaner should be circulated for a minimum of 8 hours the longer the better if time permits in order to assure thorough cleaning Filters ALPS filter All models except HighQ 2 and Spirit One Wh...

Page 11: ...ntifrogen L coolant solution will ensure a correct coolant level readout of the chiller Fill the chiller reservoir with a mixture of Antifrogen L and distilled water ratio 1 3 Do not use deionized wat...

Reviews: