  File: SG524_r3eng - April 2017 - Technical note: SG 524                                                                                                                        page   15  of  16

PM 250 

        SG 524

ALARM TROUBLESHOOTING *

Note*: the last alarm occured is permanently stored by the CPU, even in absence of main line, and 

can be verified using the Front Panel software (vers. 7.20 and over)

Code

Description

Cause/Solution

0001

Power 

Module Over 

Temperature

a) Check if the fans are working or the rear of the air is obstructed 

(heatsink fins side)

b) Check also the inlet openings for the incoming air flow.

c) Reduce ambient temperature
 If the alarm persist, contact the service

0010

Reverse Power 

Alarm

The reflected power has reached the alarm threshold. Set higher 

threshold or start the automatic tuning to deacrese the reflected power.

0011

Power Module 

Malfunction 

(PLL)

Alarm linked to internal frequency generator. The CPU turns off the power 
module automatically. After the alarm reset operation, the CPU turns 
on the power module and checks possible alarms. If the alarm persist, 
contact the service.

0100

Internal 

Safety Relay 

Malfunction

The internal safety relay could have been damaged (i.e by a shortcircuit 
on the contact)
Try to restart the unit. If the alarm persist, contact the service.

0101

Power Supply 

Undervoltage

The AC/DC power supply could have been damaged. Try to restart the 
unit. If the alarm persist, contact the service.
Note:  to reset the alarm you have to turn off the unit.

0110

AC/DC 

Power Supply 

Overtemper.

The fans of the AC/DC psu are not working properly and/or the ambient 
air condition is too hot. Let the unit cool down, run at lower power, replace 
fans if failed

0111

Internal 

Overload (M1)

The circuit M1 has reached the limit current. Try to reset and restart at 
lower power. If the alarm persist, contact the service.

1000

Internal 

Overload (M2)

The circuit M2 has reached the limit current. Try to reset and restart at 
lower power. If the alarm persist, contact the service.

1001

Open Interlock

Check interlock contact (OPEN = Alarm, closed = OK)

1010

Communication 

interrupted

Alarm engaged only when the SG is run throught the serial port and no 
valid character is received within time out

Summary of Contents for SG 524

Page 1: ...SOLID STATE MICROWAVE GENERATOR 450 W 2450 MHZ TECHNICAL NOTE issue April 2017 SG 524...

Page 2: ...Cover Rear...

Page 3: ...i membri relative alla compatibilit elettromagnetica Conformity to the Directives is assured through the application of the following standards La conformit alle Direttve assicurata tramite l applicaz...

Page 4: ...tion alone must be included into a larger process equipment and it s for stationary use Nevertheless we invite the User to respect the following Directive rules this WEEE is not a municipal waste and...

Page 5: ...Industrial scientific and medical ISM radio frequency equipment Electromagnetic disturbance characteristics Limits and methods of measurement With reference to Basic Standard EN55011 this equipment i...

Page 6: ...lay version with front panel manual commands and PLC interface to control by I O it carries also one USB standard port which allows to drive read data from the unit with a simple software like our Fro...

Page 7: ...f 9 450 Adjustable range 2 100 Max value differs at frequency band limit Output linearity Wrf 5 Forward power Displayed value vs monitor signal Output stability 1 For input fluctuations within 10 Outp...

Page 8: ...your external contactor Once switched OFF wait at least two seconds before setting it ON again The OFF state between two consecutive ON state must last two seconds minimum The following description is...

Page 9: ...LINE BREAKER 0 1 in pos 1 the unit goes in POWER ON The line switch can be freely switched off at any time POWER SET it s an analog input value and has the meaning of desired percent of the full outp...

Page 10: ...u generate a directory named program_name containing the micro code to be loaded an executable program named SendApplicationProgram bat and the loader sRecLoad exe Once unpacked the zip program enter...

Page 11: ...generator was running before entering into autotune mode The user may adjust the various parameters via three buttons and a rotating knob on the front panel as follows ESC button used to return back t...

Page 12: ...n case of heavy standing wave the generator may be overloaded with the sum of both forward and reverse power and parts of the circuit on the final power combiner as well as the coaxial socket may be d...

Page 13: ...84810 1 Preferred size other length on request max 2 m Coax Cable 7 8 with 7 16 Plugs l 1m 441384811 1 Other length on request max 4 m 3 Transition 7 16 to WR340 MWR26 COAX7 16L Aluminum construction...

Page 14: ...gnals in the range 0 10Vdc pin 2 and pin 3 ground on pin 15 and 16 respectively Forward and Reverse Power Monitor CONN 2 LOCATION Forward Power Wrf 1 Vmon 45 W Pin no 2 return on pin 15 Reverse Power...

Page 15: ...for the protocol implemented Each SG 524 generator must be associated to an address the two digital rotating switches must be used the left switch sets the tens the right sets the units Depending fro...

Page 16: ...1 1 Plug for CONN 1 included on Plug Set MOCPC03 3 1 Connector Sub D type 25 pin female CONN 2 4 1 AC DC power supply 120 240V 32 V out PLRSP200032 5 1 Anybus CC interface Profibus Modbus PLAB62xxx S...

Page 17: ...alarm persist contact the service 0100 Internal Safety Relay Malfunction The internal safety relay could have been damaged i e by a shortcircuit on the contact Try to restart the unit If the alarm per...

Page 18: ...ET524 contains plugs for Conn 1 and 5 only Coax Plug right angle 7 16 M F p n MWCOAX7 16L Coax cable length 1 m 1 2 p n 441384810 1M0 Same as above 2 m p n 441384810 2M0 Cable set for input line and s...

Reviews: