First version
Rev 1.00
22 Sep 2016
Helios User Manual
22
64 00 - min exposure time (ms). Hex value 0x0064, decimal 100 (ms)
E0 2E - max power (W). Hex value 0x2EE0, decimal 12000 (W) or 12 kW
64 00 - min power (W). Hex value 0x0064, decimal 100 (W)
3C 00 - max allowed temperature of Helios. Hex value 0x003C, decimal 60 degrees C.
00 00 - spare (2 bytes)
00 00 00 00 - last power measurement (mW)
00 00 00 00 - last time measurement (μs)
00 00 00 00 - last energy measurement (mJ)
81 00 00 00 - current power (mW). Hex value 0x00000081, decimal 129 mW (note, the current power can be
negative, in which case it will be displayed as 2s-compliment, for example 27 FF FF FF, Hex value 0xFFFFFF27,
decimal -217 mW)
1E 00 - temperature of Helios. Hex value 0x1E, decimal 30 degrees C
01 00 - current energy scale is index 0x0001 (10kJ scale) - other scales are 2 (1kJ scale) and 3 (100J scale)
10 27 00 00 - max energy in current scale (J). Hex value 0x00002710, decimal 10000 (J) or 10kJ
00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 00 - spare bytes
For examples of how to send commands and receive status using PROFINET see Chapter 11 below.
For examples of how to send commands and receive status using EtherNet-IP see Chapter 10 below.