CONEX-LDS
Controller GUI Manual
Choosing “Yes” will save the parameters in memory and “No” will cancel the
operation.
Set Volatile
Writes to volatile memory and is lost upon reboot.
4.6.2
Dark current calibration
Update Button
Updates the offsets. Follow the pop up messages to complete the process.
NOTE
This function minimizes the effects of internal reflections that reach the sensor
and impact the measurement signal. The laser should be ON for a least 10
minutes before performing this calibration. A non-reflective shutter should be
inserted as close as possible to the autocollimator beam output.
This calibration is performed on each CONEX-LDS at the factory and can be
performed on site. Be familiar with the process before attempting to proceed,
since it could be source of measurement errors, if not properly done.
4.6.3
Low pass filter
Set volatile Button
Applies the new value of the low pass filter frequency.
Set permanent Button
Saves the new value of the low pass filter frequency in the memory of the CONEX-
LDS. The controller will need to b in Configuration state, in order for the changes to
take place.
19
EDH0331En1032 — 12/18