Installation
Series 358 Micro-Ion Controller Instruction Manual - 358013 - Rev. B
63
Sy
stem C
o
mponents
Ini
tial
Setup
Installation
Before You
Begin
4.14
Connecting Process
Control Relays
For instructions for setting up this module, see
The process control connector is embossed with letters identifying corner
pins. Table 4-3 shows the letters designating the 3 pins assigned to each of
the 6 setpoint channels.
Figure 4-8
Process Control Output Connector
Table 4-3
Process Control Output Connector Pin Assignments
Process Control Channel
Micro-Ion Gauge
Convectron Gauge A
Convectron Gauge B
1
2
3
4
5
6
Common (or Pole)
W
H
M
C
J
S
Normally Closed (NC)
P
A
U
K
B
X
Normally Open (NO)
T
D
R
F
E
V
CHASSIS GND – PIN L
NO CONNECTION – PIN N
A mating connector is supplied in the hardware kit.
CAUTION
Failure to check system programming before switching
to automatic operation can cause measurement error.
To avoid measurement error due to inaccurate output signals,
carefully check the system programming before switching to
automatic operation.
Summary of Contents for Granville-Phillips 358 series
Page 2: ......
Page 8: ...Table of Contents 8 Series 358 Micro Ion Controller Instruction Manual 358013 Rev B...
Page 26: ...Chapter 2 26 Series 358 Micro Ion Controller Instruction Manual 358013 Rev B...
Page 48: ...Chapter 3 48 Series 358 Micro Ion Controller Instruction Manual 358013 Rev B...
Page 94: ...Chapter 5 94 Series 358 Micro Ion Controller Instruction Manual 358013 Rev B...
Page 104: ...104 Series 358 Micro Ion Controller Instruction Manual 358013 Rev B...
Page 116: ...Chapter 8 116 Series 358 Micro Ion Controller Instruction Manual 358013 Rev B...
Page 120: ...120 Series 358 Micro Ion Controller Instruction Manual 358013 Rev B...
Page 121: ......