Fittings
Chapter Two: Installation
24
Fittings
The transducer port will easily carry the weight of the transducer. The following is a list of
available fittings:
•
½” diameter (12.7 mm) tubulation
•
NW16-KF
•
Swagelok
®
8-VCR
®
(female)
•
Swagelok 8-VCO
®
(female)
•
1.33 inch (33.8 mm) O.D. CF
Caution
MKS
does not warranty
the 622B, 623H, or 626D transducers when
single or double metal ferrule compression-type vacuum fittings
(shown in Figure 3, page 24) are used because damage will occur to
the transducer when improper tightening procedures are followed.
Figure 3: Metal Single and Double Ferrule Compression-type Vacuum Fittings
Single Ferrule
Double Ferrule