Calibration Procedure
Your CobraTurn® T-260 comes calibrated from the factory, but if you
suspect the unit's RPM readings are off, follow the procedure below to
re-calibrate the unit:
1. Make sure the unit is off.
2. Press and
hold down
the direction button, then press the power
button.
3. Next release the power button and then the direction button.
4. If done correctly, "CAL" will be displayed as shown.
5. Begin the calibration by pressing the start button (knob / push button).
Calibration will take about 20 minutes as the table turns and records
RPM and time data.
6. After the data is recorded, the letters "SLo" will be flashing on the
display, s
lowly turn
the speed knob until the turntable barely begins to
rotate.
7. Finally press the stop button (knob / push button) to finish and save the
calibration.
Weld Ground 5 AMP Fuse
Your CobraTurn® T-260 is outfitted with a 5 AMP fuse to protect the
internal electronics in the event of improper grounding. Replace with
5AMP 250vac 3AG fuse.
Circuit Breaker
May be tripped in the event of an input power surge or if excess
voltage is applied. Breaker switch must be reset.