6875 hana Table
T e c h n i c a l S e r v i c e s D e p a r t m e n t
Preventative Maintenance Checklist Phone: 800-777-4674 PST Fax: 510-429-8324
50603-06 Rev. A ECN# 10016 Effective Date 3/9/2010 Page 3 of 4
3.3 ( ) ( ) Attach Right & Left Femur Lift assemblies, and connect their electrical cords.
(Confirm green “LEFT LIFT CONNECTED” and “RIGHT LIFT CONNECTED” lights are on)
Comments:__________________________________________________________________________
3.4 ( ) ( ) Attach Right Leg Spar & Left Leg Spar.
a. Verify locking knob locks the spar to the table with ease.
b. Verify Traction unit slides smoothly onto the rail and locking knob locks & unlocks the Traction unit.
c. Verify Rotational Lock knob is operational and its free of any cracks and/or any other abnormality.
d. Verify the spar brake lock handle is operational and the spar does not drift down when locked.
e. Verify Fine Traction when crank has a little bit of tension (If it’s too smooth, please note or call
Technical Services).
Comments: _________________________________________________________________________
4. Control Panel check.
4.1 ( ) ( ) Power Status Lights; observe the table Power Status lights located on the Control Panel. Confirm green
“AC Power” light is on. If red “Battery Check” light is on, charge battery for a minimum of 3 hours
(leave Power Cord plugged in and main “ON/OFF” Power Switch “ON”).
a.
Press and hold "HEIGHT UP" / “HEIGHT DOWN" button (run to its limit).
b.
Press and hold "REV. TREN"/ "TREN" button (run to its limit).
c.
Press and hold "LEFT LATERAL “/ “RIGHT LATERAL ROLL” button (run to its limit).
d
. “RETURN TO LEVEL (run table to max height up, rev tren, and lateral roll right. Press and hold
Return to Level button, verify table goes back to level (-+2deg) and to the lowest position.
e.
Switch the selector knob to “Left Femur Lift”, press “UP”/”DOWN” on the foot pedal. Confirm
Femur Lift goes up/down to its limit.
f.
Switch the selector knob to “Right Femur Lift”, press “UP”/”DOWN” on the foot pedal. Confirm
Femur Lift goes up/down to its limit.
g.
Press “Emergency Button”. Confirm all indicators for “AC POWER”, “BATTERY OK”,
“BATTERY POWER”, and “BATTERY CHARGE” will not illuminate except for indicators, “Foot
Pedal Connected, “Left” / “Right” Lift Connected, and “Left / Right” Femur lift selector.
Comments: _________________________________________________________________________
5. Hand Pendant check
.
5.1 ( ) ( ) Confirm green “AC Power” light is on and green “Battery Status” light is on.
(
Repeat steps 4.1, a to d
)
Comments:
_________________________________________________________________________
6. Battery Power Check (disconnect power cord from AC source. Main power switch indicator will dim).
6.1 ( ) ( ) Confirm the green, “Battery OK” lights is ON at the Control Panel and Hand Pendant are illuminated. If
not, refer to the Battery System section of the service manual.
(
Repeat steps 4.1, a to g for this section)
Comments: _________________________________________________________________________