Specifications
Description
Operating Table VACB-3006B
Electrically-powered
Elevation
range
Highest
1060 mm
Lowest
704 mm
Trendelenburg
angle
Head down
20°
Head up
20°
Lateral tilt
angle
Right down
20°
Left up
20°
Return to level
Trendelenburg / Lateral tilt
Brake
Lock / Unlock
Control
devices
Control unit
NORMAL, REV, Trendelenburg, Lateral tilt, Elevation, Return to level,
Brake, E button, On/Off
Control handle
Power on, Slide unlock
Foot Switch
Elevation, Trendelenburg, Lateral tilt :
Note 6
Auxiliary switch
Trendelenburg, Lateral tilt, Brake, Slide unlock
Manual
Max. sliding
Length
710 mm
Width
200 mm
Others
Emergency brake release handle, Auxiliary brake
Electrical rating
Classification per IEC 60601-1
Class 1 (internal power source device :
Note 1
) / TypeB / IPX4
Rated supply voltage
AC100-240 V
Rated supply frequency
50-60 Hz
Battery power supply voltage
DC24 V
:
6 V 10 Ah x 4
Power input
400 VA
Operating voltage
DC5 V, 24 V
Duty cycle
3 min on, 7 min off :
Note 2
Others
Recovery from defibrillator is within five seconds.
This product conforms to EMC Standard IEC 60601-1-2
2014
Dimension
Tabletop dimension
2095 mm (L) x 500 mm (W) :
Note 3
Base dimension
1244 mm (L) x 486 mm (W)
Weight
385 kg
Maximum permissible load
Lift: 225 kg / Trendelenburg, Lateral tilt: 135 kg
Installing
environment
Ambient temperature
10°C - 40°C:
Note
7
Relative humidity
30% - 75%:
Note
7
Atmospheric pressure
700 hPa - 1060 hPa:
Note
7
Other
Allowable altitude for use is
3
000 m or lower
:
Note 7
Transporting and
storing condition
Ambient temperature
-10°C - 50°C :
Note 4
Relative humidity
10% - 85% (No condensation):
Note 4
Atmospheric pressure
700 hPa - 1060 hPa :
Note 4
Transitable height and width
height 10 mm
width 80 mm
Useful life
When that specified maintenance and appropriate storage are done
10 years :
Note 5
Accessories:
Note 6
Side Rail Adaptor II, Long for Carbon Fiber Board (R1/R2)
Side Rail Adaptor II, Short for Carbon Fiber Board (R1/R2)
Radiolucent Head Extension for VACB with Pad
Note 1:
when the battery power is used
Note 2:
consecutive pressing the switch of the control unit.
Note 3:
excluding the side rail
Note 4:
Company standard
Note 5:
based on the self-certification (company data)
Note 6:
Option
Note 7
:IEC 60601-1
2012
, Medical electrical equipment - Part1: General requirements for safety
Specifications
Summary of Contents for VACB-3006B
Page 2: ......
Page 4: ......
Page 5: ...1 Introduction 1 Introduction ...
Page 6: ...1 Introduction ...
Page 8: ...1 Introduction ...
Page 9: ...2 Before use 2 Before use ...
Page 17: ...3 Component Identification 3 Component Identification ...
Page 43: ...5 Maintenance and inspection 5 Maintenance and inspection ...
Page 46: ...42 5 Maintenance and inspection ...
Page 47: ...6 Specifications 6 Specifications ...
Page 50: ...46 6 Specifications ...
Page 51: ...7 When the control unit can not be used 7 When the control unit can not be used ...
Page 56: ...52 8 Before contacting your distributor for repair ...
Page 57: ...Appendix ...