![Mizuho VACB-3006B Operator'S Manual Download Page 12](http://html1.mh-extra.com/html/mizuho/vacb-3006b/vacb-3006b_operators-manual_1812837012.webp)
8
2
Before use
When using with a C-arm X-ray machine
- Read any documents or instruction manuals included with the C-arm X-ray machine to fully
understand any influence that the machine may have on the operation table. The transmission of
X-rays may decrease because of the carbon-made tabletop.
- Make sure the C-arm X-ray machine does not come in contact with the table. Improper installation
may obstruct operations or cause a malfunction.
Allowable load
- Do not exceed the allowable load*.
The table may become inoperable and cause malfunction.
*Lift: 225 kg / Trendelenburg, Lateral tilt: 135 kg
Preventive maintenance and inspection
- Make sure to perform inspection before and after use, and appropriate preventive maintenance
inspection.
The table may not survive its useful life depends on number of years and frequency of use.
- For preventive maintenance inspection, please contact your distributor or us directly.
Cardiac defibrillator
Consult manufacturer’s instructions when using high frequency surgical equipment, cardiac
defibrillator and cardiac defibrillator monitors.
Antistatic pathway
When an antistatic pathway is required, the table has to be used on an antistatic floor.
WARNING
Prohibition of disassembling and modification
- Do not attempt to disassemble or modify the table. It may cause a malfunction.
When used together with other medical electrical equipment
- Before using together with other equipment, make sure to check that it is not affected by
electromagnetic interference.
- Electromagnetic interference from other medical electrical equipment may cause malfunction.
When performing CPR, slide the tabletop toward the leg side as far as it will go.
Improper body position may make the table unstable and cause the patient to get injured.
Others
Summary of Contents for VACB-3006B
Page 2: ......
Page 4: ......
Page 5: ...1 Introduction 1 Introduction ...
Page 6: ...1 Introduction ...
Page 8: ...1 Introduction ...
Page 9: ...2 Before use 2 Before use ...
Page 17: ...3 Component Identification 3 Component Identification ...
Page 43: ...5 Maintenance and inspection 5 Maintenance and inspection ...
Page 46: ...42 5 Maintenance and inspection ...
Page 47: ...6 Specifications 6 Specifications ...
Page 50: ...46 6 Specifications ...
Page 51: ...7 When the control unit can not be used 7 When the control unit can not be used ...
Page 56: ...52 8 Before contacting your distributor for repair ...
Page 57: ...Appendix ...