71
8
Specification
Transitable height and width
Height : 10 mm (0.93 in) / Width : 80 mm (3.15 in)
Operating environment
Temperature
10 to 40°C : Note 8
Humidity
30 to 75% : Note 8
Atmospheric pressure
700 to 1,060 hPa : Note 8
Others
Allowable altitude for use is 3,000 m or lower. : Note 8
Transportation
and storage
Temperature
-10 to 50°C : Note 9
Humidity
10 to 85% (without moisture condensation) : Note 9
Atmospheric pressure
700 to 1,060 hPa : Note 9
Service life
Under the specified maintenance and proper storage, 10 years : Note 10
Note 1: Optional
Note 2: When the battery power is used
Note 3: Tables continuously operating possible time and halting time.
Note 4: Excluding the side rail
Note 5: Rough dimension
Note 6: Total of the patient and accessories
Note 7: 290 to 340 kg (640 to 740 lbs) when the table top height is 400 to 550 mm (15.75 to 21.65 in).
(Lifting only)
Note 8: IEC 60601-1, Medical electrical equipment - Part1: General requirements for safety
Note 9: Company standard (in case that appropriate maintenance and inspection is done)
Note 10: Based on Mizuho's own validation data