69
8
Specification
Rating
Classification as per 60601-1
Class I device (internal power source device : Note 3)
Rated supply voltage
AC 100 - 240 V
Rated supply frequency
50-60 Hz
Battery power
DC 24 V
Power input
700 VA
Operating voltage
24 V
Duty cycle
3 min on, 7 min off : Note 4
Others
Recovery from defibrillator is within five seconds.
Conformity to EMC Standard IEC 60601-1-2: 2018
Applied part
B type applied part
Head plate / Auxiliary back plate / Back plate / Seat plate /
Leg plate
Defibrillation-proof applied part
Mattresses (Head plate / Auxiliary back plate / Back plate /
Seat plate / Leg plate)
Water resistant class
IPX4
Dimension
Tabletop
2,100 mm (82.68 in) (L) x 500 mm (19.69 in) (W) : Note 5
Base
1,335 mm (52.26 in) (L) x 500 mm (19.69 in) (W) : Note 6
Weight
365 kg (805 lbs)
Allowable patient body weight: Note 7
360 kg (800 lbs) : Note 8
Transitable height and width
Height : 10 mm (0.39 in) / Width : 80 mm (3.15 in)
Operating environment
Temperature
10 to 40°C : Note 9
Humidity
30 to 75% : Note 9
Atmospheric pressure
700 to 1,060 hPa : Note 9
Others
Allowable altitude for use is 3,000 m or lower. : Note 9
Transportation
and storage
Temperature
-10 to 50°C : Note 10
Humidity
10 to 85% (without moisture condensation) : Note 10
Atmospheric pressure
700 to 1,060 hPa : Note 10
Service life
Under the specified maintenance and proper storage, 10 years : Note 11
Note 1: from the center position
Note 2: Optional
Note 3: When the battery power is used
Note 4: Tables continuously operating possible time and halting time.
Note 5: Excluding the side rail
Note 6: Rough dimension
Note 7: Total of the patient and accessories
Note 8: 290 to 340 kg (640 to 740 lbs) when the table top height is 450 to 600 mm (17.72 to 23.62 in).
(Lifting only)
Note 9: IEC 60601-1, Medical electrical equipment - Part1: General requirements for safety
Note 10: Company standard (in case that appropriate maintenance and inspection is done)
Note 11: Based on Mizuho's own validation data