10
2
Before use
Others
CAUTION
When used together with other medical equipment or accessories
- Before using another devices or accessories, read carefully the instruction manual of the device and
make sure that the operating table will not be affected adversely. Before fitting an accessory of third
party company, make sure to contact your distributor or us. Some accessories cannot be fitted.
- When using the operating table, make sure to check the positions of other equipment.
They may come in contact with each other during the operation, and may get damaged.
- For hygiene, be sure to use sterile drapes on the areas on this product where the patient comes
into contact with it.
Disinfection
- The following procedure should be followed when disinfecting an operating table.
1. Power off and remove the power cord from medical grade wall outlet.
2. Remove all table pads from the operating table.
3. Apply a proper quantity of disinfectant on a clean and lint-free cloth, and wipe the top and
sides and bottom of the pads with the cloth. A proper quantity of disinfectant means a quantity
such that the cloth with disinfectant does not dry when wiping.
4. Disinfect the table board of the table top and the surface of the side rail in the same way.
5. Wipe all disinfected parts with dry, clean and lint-free cloth within 15 minutes after disinfecting.
Usable disinfectant solutions which we have already confirmed as compatible:
1 Sodium hypochlorite 0.1% (halogen containing compound)
2 Sodium thiosulfate
3 Chlorhexidine (chlorhexidine gluconate 0.5%)
4 Benzalkonium chloride (invert soap 10%)
5 Povidone iodine
6 Ethanol 80%
7 Hydrogen peroxide
8 Saline (0.9%)
9 Isopropyl alcohol (IPA)
Move and transportation
- Do not move the table while a patient is on it.
- Follow the procedure below for moving the table.
* When moving the operating table, disinfect the whole operating table in order to prevent infection.
1. Power off and remove the power cord from medical grade wall outlet.
2. Check if handles and levers are in the fixed position, and each section is fixed firmly.
3. Unlock brake to transfer the table.
* Do not move the operating table by pulling on the head plate or leg plate.
- Transportation should be carried out with the following conditions met.
1. Disinfect the whole operating table before transportation.
2. Take measures against falling down, like moving down the table to the lowest position.
3. Actuate a brake.
4. Use cushioning materials to prevent damage during transportation.
5. Protect against dust, wind and rain.
Disposal
- In accordance with the European Union Waste Electrical and Electronic Equipment (WEEE)
Directive, all electrical components and batteries must be disposed of in accordance with local
regulations. Please contact your local distributor for proper disposal.
Summary of Contents for MOT-VS600
Page 2: ......
Page 4: ......
Page 5: ...1 Introduction 1 Introduction ...
Page 6: ...2 1 Introduction ...
Page 10: ...6 1 Introduction ...
Page 11: ...2 Before use 2 Before use ...
Page 19: ...3 Component identification 3 Component identification ...
Page 24: ...20 3 Component identification ...
Page 60: ...56 4 Operation ...
Page 61: ...5 Maintenance and inspection 5 Maintenance and inspection ...
Page 64: ...60 5 Maintenance and inspection ...
Page 65: ...6 Specifications 6 Specifications ...
Page 69: ...7 Troubleshooting 7 Troubleshooting ...
Page 71: ...8 Before contacting for repairs 8 Before contacting for repairs ...