Mizuho MOT-5602BW Operator'S Manual Download Page 47

45

7

T

roubleshooting

„

z

Use the emergency brake release dial to release the 

brake

WARNING

Do not tilt the tabletop while the emergency brake 

release dial is in the "UNLOCK" position.

The patient may fall from the operating table.

In case of electrical trouble, the operating table can be moved by using the 

emergency brake release dial.

Follow the procedure below to release the brake.

1. 

Turn the emergency brake release dial toward "UNLOCK".

The brake is released.

Turn the dial

WARNING

To move the operating table after you released 

the brake, activate the brake again with the 

following procedure. If the operating table is 

moved with the brake released, the patient may 

fall from the operating table.

1.  Return the emergency brake release dial to 

"LOCK" direction.

Turn the dial

2.  Press the 

ENERGIZE

 and 

BRAKE UNLOCK

 on the control unit for 

1 second or more.

3.  Hold 

ENERGIZE

 and press 

TABLE UP

 on the control unit. 

The operating table is fixed.

Summary of Contents for MOT-5602BW

Page 1: ...rposes may result in damage or injury The operator and the person in charge of the maintenance of this operating table must read this operator s manual thoroughly and understand the contents before operating inspecting adjusting and maintaining it Keep this manual for reference in a place where is readily accessible MES CK18 450 10EN Ver 2 2019 5 ...

Page 2: ...le 23 4 5 Tilting the tabletop laterally 24 4 6 Trendelenburg 25 4 7 Tilting the back plate 26 4 8 Changing the tabletop height 27 4 9 Sliding the tabletop 28 4 10 Flexing or reflexing the tabletop 29 4 11 Returning to level 31 4 12 Adjusting the head plate 32 4 13 Adjusting the leg plate 35 5 Maintenance and inspection 39 5 1 Inspection before and after use 39 5 2 Periodic replacement parts 41 5 ...

Page 3: ...otes additional information on the product s functions The warning and caution notices on this manual relating to operating inspecting and maintaining apply to the intended use surgical operations of this product If the product is used for purposes other than surgery the user is responsible in regard to safety for performing operations inspections and repairs which are not contained in this manual...

Page 4: ...2 1 Introduction 1 3 Accessories Standard components and accessories Control unit Operators manual Mattresses for head back and hips and legs Main unit ...

Page 5: ...g in injury CAUTION Before lowering the table or placing it in a reverse Trendelenburg position check if there are any devices under the leg plates If the leg plates come in contact with devices that are to be subjected to excessive force the leg plate insertion shaft may be damaged CAUTION When transferring a patient from a surgical bed Fig a or changing a patient s body position Fig b do not app...

Page 6: ... any objects in the place where the power cord is to be unplugged from the medical grade outlet which would obstruct it from being unplugged Power cord 5 Installation of the operating table WARNING Do not install the operating table on an uneven floor Do not place a pad under the base for raising the operating table The operating table may tip over resulting in injury Uneven floor Pad Patient s po...

Page 7: ...uency surgical equipment and or a cardiac defibrillator refer to their operator s manual provided by the manufacturers Improper usage may cause the operator and the patient to get burned and or devices to malfunction Patient position When using the tabletop or accessories to secure a patient s body position always observe the patient s condition Neuroparalysis may occur to the patient Allowable lo...

Page 8: ...th it Cleaning and disinfection After using the operating table make sure to follow the steps below to clean up and disinfect the operating table 1 Turn off the power and disconnect the power cord from the medical grade outlet 2 Detach all the mattresses from the operating table 3 Use a lint free cloth soaked with proper volume of disinfectant to wipe off the upper side and back side of the mattre...

Page 9: ...y 3 Unlock the brakes and move the operating table The operating table should be transported with the following conditions met Disinfect the entire operating table before transporting it Take measures to prevent it from tipping over such as lowering the tabletop to the bottom position Actuate the brake Suitably position cushioning on the product to prevent it from getting damaged during transport ...

Page 10: ... table is labeled at the locations shown as below Before use make sure to understand the contents of the labels Warning and Caution labels 1 C655650 2 C655608 3 C653624 6 C655740 7 C655803 9 C653614 4 C655732 5 C655649 8 C657333 1 2 9 8 7 4 3 5 4 3 6 ...

Page 11: ...9 2 Safety precaution Other labels 1 2 10 C600516 11 C655646 12 C600020 13 C655648 14 C653515 16 C600559 15 C653516 17 C653513 18 C630006 19 C655683 17 10 19 16 11 10 13 14 15 12 11 12 18 ...

Page 12: ...2 20 C653620 21 C655691 ASIA 22 C655692 ASIA 23 C646067 ASIA 24 PIN 630D8M3 25 PIN 630D8M4 26 PIN 562G6M2 27 PIN 50005M2 C655689 EU or C655686 US or C655690 EU or C655687 US or C646070 EU or C646073 US or 20 20 20 20 24 23 27 26 25 22 21 20 ...

Page 13: ...s manual 6 9 13 18 General prohibition sign 6 General mandatory action sign 6 Emergency stop 9 Equalization terminal 17 Indicates AC power supply 18 Enclosure Class Splash proof 18 Serial number 18 Product Number 18 Waste disposal information 18 Authorized representative in the European Community 18 Manufacturer 18 24 25 26 27 Type B 20 Defibrillation proof Type B applied part 24 25 26 27 ...

Page 14: ...s Leg plate 1 Control unit Waist plate Back plate Emergency stop switch Head plate 1 Leg plate flexing lever Leg plate fixing knob Emergency brake release dial Caster Head plate fixing knob Side rail Head plate flexing lever Side frame Sliding cylinder Column Base Control unit connector Brake Fig A 1 The head plate and the leg plate are detachable 2 The foot switch is optional ...

Page 15: ...lenburg Head down Lateral tilt Right down Back plate bending Down Lift Down Slide To leg Reflex Level Power lamp Power ON OFF switch Reverse Trendelenburg Head up NOTE The switches other than the E switch continue to function while being pressed If a switch is continuously pressed for 3 minutes or more its operation is halted The halt condition continues for about 7 minutes Switches other than the...

Page 16: ... 5 m or more 2 5 m or more 2 5 m or more Installing the operating table CAUTION When this product is used with the power cord disconnected use the equipotential wire to ground the equipotential terminal to the medical grounding terminals Prepare the equipotential wire yourself Equipotential wire Equipotential terminal 1 Move the operating table to a flat area ...

Page 17: ...eptacle properly Insert the connector Detaching the control unit When you replace the control unit detach the connector from the receptacle 1 Turn the connector ring in the direction of the arrow until it stops 2 After the ring stopped pull out the connector upward 1 Turn the ring 2 Pull out the connector ...

Page 18: ...tery naturally discharges itself when it is not being used and is being stored Make sure to charge the battery before use While the battery is being charged you can operate the operating table by using ON OFF on the control unit If the battery lamp on the control unit blinks after usage of the operating table or the battery indicator on the lower part of the operating table lights up only red whil...

Page 19: ... function are functioning properly 1 Connect the power connector of the product and the medical grade outlet with the power cord 2 Press ON OFF of the control unit 3 Press ENERGIZE and one of the function switches on the control unit at the same time If two lines of green of the battery indicator light up the control unit is functioning properly Control unit is normal Green ...

Page 20: ...TO LOCK BRAKES LEVEL BRAKE UNLOCK REFLEX FLEX SLIDE HEAD SLIDE FOOT TABLE DOWN BACK UP BACK DOWN TILT LEFT TILT RIGHT REV TREND POWER BATT PILOT MOT 5602BW TREND NOTE In an emergency or when turning off the power completely disconnect the power cord from the medical grade outlet If you press ON OFF on the control unit before turning on the power switch on the base the power of the battery is turne...

Page 21: ...UNLOCK REFLEX FLEX SLIDE HEAD SLIDE FOOT TABLE DOWN BACK UP BACK DOWN TILT LEFT TILT RIGHT REV TREND POWER BATT PILOT MOT 5602BW TREND 2 Turn off the power switch on the base The power switch green and battery indicator go out and the power is turned off NOTE If you turn off the power switch on the base before pressing ON OFF on the control unit the power of the battery is turned on ...

Page 22: ...RESS TO LOCK BRAKES LEVEL BRAKE UNLOCK REFLEX FLEX SLIDE HEAD SLIDE FOOT TABLE DOWN BACK UP BACK DOWN TILT LEFT TILT RIGHT REV TREND POWER BATT PILOT MOT 5602BW TREND NOTE In an emergency or when turning off the power completely disconnect the power cord from the medical grade outlet During use with the battery power the power is automatically turned off if no operation is performed for 2 hours or...

Page 23: ...off The power lamp green and battery lamp orange on the control unit and battery indicator go out and the power is turned off TABLE UP ENERGIZE ON OFF PRESS TO LOCK BRAKES LEVEL BRAKE UNLOCK REFLEX FLEX SLIDE HEAD SLIDE FOOT TABLE DOWN BACK UP BACK DOWN TILT LEFT TILT RIGHT REV TREND POWER BATT PILOT MOT 5602BW TREND ...

Page 24: ...switch The buzzer sounds and the operating table stops Canceling operations 1 After the operating table stops turn the emergency stop switch in the direction of the arrow to cancel the emergency stop switch The buzzer stops sounding CAUTION To reset the operating table to the original position in an emergency where for example an operator s hand is caught in a gap of the operating table press the ...

Page 25: ...To move the operating table unfix the operating table WARNING Do not unfix the operating table with a patient on it The patient may fall from the operating table 1 Hold down ENERGIZE and press BRAKE UNLOCK for one second or more The brake is released for unfixing the operating table TABLE UP ENERGIZE ON OFF PRESS TO LOCK BRAKES LEVEL BRAKE UNLOCK REFLEX FLEX SLIDE HEAD SLIDE FOOT TABLE DOWN BACK U...

Page 26: ... to the right in the view from the head side TABLE UP ENERGIZE ON OFF PRESS TO LOCK BRAKES LEVEL BRAKE UNLOCK REFLEX FLEX SLIDE HEAD SLIDE FOOT TABLE DOWN BACK UP BACK DOWN TILT LEFT TILT RIGHT REV TREND POWER BATT PILOT MOT 5602BW TREND Right down NOTE The maximum angle achieved in the right down and left down position is 20 to the level position 20 20 20 20 If the slide cylinder is tilted close ...

Page 27: ...TREND POWER BATT PILOT MOT 5602BW TREND Head up position Trendelenburg Head down 1 Hold down ENERGIZE and press TREND The tabletop moves to the head down position TABLE UP ENERGIZE ON OFF PRESS TO LOCK BRAKES LEVEL BRAKE UNLOCK REFLEX FLEX SLIDE HEAD SLIDE FOOT TABLE DOWN BACK UP BACK DOWN TILT LEFT TILT RIGHT REV TREND POWER BATT PILOT MOT 5602BW TREND Head down position NOTE The maximum angle of...

Page 28: ... BRAKE UNLOCK REFLEX FLEX SLIDE HEAD SLIDE FOOT TABLE DOWN BACK UP BACK DOWN TILT LEFT TILT RIGHT REV TREND POWER BATT PILOT MOT 5602BW TREND 40 NOTE The maximum angle achieved in the back plate up position is 90 to the level position The maximum angle achieved in the back plate down position is 40 to the level position When the tabletop center point is slid in the foot direction beyond the standa...

Page 29: ...he tabletop 1 Hold down ENERGIZE and press TABLE DOWN The tabletop moves down TABLE UP ENERGIZE ON OFF PRESS TO LOCK BRAKES LEVEL BRAKE UNLOCK REFLEX FLEX SLIDE HEAD SLIDE FOOT TABLE DOWN BACK UP BACK DOWN TILT LEFT TILT RIGHT REV TREND POWER BATT PILOT MOT 5602BW TREND 520 mm 20 5 in Move down NOTE The maximum height from the floor to the tabletop upper surface is 1000 mm 39 4 in The minimum heig...

Page 30: ...AKES LEVEL BRAKE UNLOCK REFLEX FLEX SLIDE HEAD SLIDE FOOT TABLE DOWN BACK UP BACK DOWN TILT LEFT TILT RIGHT REV TREND POWER BATT PILOT MOT 5602BW TREND 220 mm 8 7 in Head direction NOTE Maximum travel in sliding from the center position of the tabletop is as follows Foot direction 310 mm 12 2 in Head direction 220 mm 8 7 in If the back plate is lower than the level position the tabletop cannot be ...

Page 31: ...ENERGIZE and press FLEX The back plate flexes downward and the waist plate turns to the Trendelenburg head up position TABLE UP ENERGIZE ON OFF PRESS TO LOCK BRAKES LEVEL BRAKE UNLOCK REFLEX FLEX SLIDE HEAD SLIDE FOOT TABLE DOWN BACK UP BACK DOWN TILT LEFT TILT RIGHT REV TREND POWER BATT PILOT MOT 5602BW TREND Center up NOTE In the position where the tabletop center is in the leg direction beyond ...

Page 32: ...get damaged 1 Hold down ENERGIZE and press REFLEX The back plate reflexes upward and the waist plate turns to the reverse Trendelenburg head down position TABLE UP ENERGIZE ON OFF PRESS TO LOCK BRAKES LEVEL BRAKE UNLOCK REFLEX FLEX SLIDE HEAD SLIDE FOOT TABLE DOWN BACK UP BACK DOWN TILT LEFT TILT RIGHT REV TREND POWER BATT PILOT MOT 5602BW TREND Center down ...

Page 33: ...o the level position after Trendelenburg lateral tilting back plate tilting and flexing are worked TABLE UP ENERGIZE ON OFF PRESS TO LOCK BRAKES LEVEL BRAKE UNLOCK REFLEX FLEX SLIDE HEAD SLIDE FOOT TABLE DOWN BACK UP BACK DOWN TILT LEFT TILT RIGHT REV TREND POWER BATT PILOT MOT 5602BW TREND NOTE Raising sliding and braking do not function ...

Page 34: ...NING Make sure to tighten the head plate fixing knob securely If the head plate moves the patient may get injured The head plate weighs 7 kg 15 lbs Pay special attention when handling it It may drop and cause damage or injury 1 Pull the head plate flexing lever in the head direction The head plate can be flexed upward and downward 2 Flex the head plate and then release the lever The head plate is ...

Page 35: ...kg 15 lbs Pay special attention when handling it It may drop and cause damage or injury 1 Loosen the two head plate fixing knobs located on the lower side of the back plate 2 Hold the both sides of the head plate firmly and pull it straight Head plate Head plate fixing knob ...

Page 36: ...y 1 Hold the both sides of the head plate firmly and align the insertion shaft of the head plate with the reception hole in the back plate and insert 2 After checking that the head plate is completely inserted tighten the two head plate fixing knobs located on the lower side of the back plate Head plate Head plate fixing knob CAUTION Insert the head plate into the back plate If the operating table...

Page 37: ... with the base and get damaged Do not move the tabletop to the head direction from the standard center position with the leg plates flexed The leg plates may get damaged Leg Plate Do not move down the tabletop if the tabletop center point is in the foot side beyond the standard center position The tips of the leg plates may come in contact with the base and get damaged 1 Pull the leg plate flexing...

Page 38: ...ixing knob one revolution and a half and loosen the leg plate fixing knob 2 Pull the leg plates outward 3 At the setting position tighten the leg plate fixing knob to fix Tight Loose Leg plate fixing knob Leg plate WARNING Make sure to lock the leg plate fixing knob If the leg plates move the patient may get injured ...

Page 39: ...ighs 8 kg 18 lbs each Pay special attention when handling it It may drop and cause damage or injury 1 Hold the ends of the leg plate at the leg side 2 Turn the leg plate fixing knob 3 Hold and pull up the both ends of the leg plate Leg plate Leg plate fixing knob ...

Page 40: ... leg plate clutch 2 While holding the ends of the leg plate at the leg side turn the leg plate fixing knob until the leg plate clutch engages with the leg plate shaft Leg plate clutch Leg plate Leg plate fixing knob WARNING Make sure to lock the leg plate fixing knob If the leg plates move the patient may get injured ...

Page 41: ...o for repairs Otherwise it may cause problems during surgery Inspect the items below If there is any problem request your distributor or Mizuho for repair 1 5 8 4 2 3 7 6 1 Mattresses z z Before use Check all the mattresses for any damage z z After use Check all the mattresses for any damage or dirt 2 Backlash of the tabletop z z Before use Check all the table plates for any backlash when jiggling...

Page 42: ...tions are working properly 5 Oil leakage z z Before and after use Check the floor or the base surface for any hydraulic oil 6 Extra switch z z Before use Press the extra switch to see if all functions are working properly 7 Power switch z z Before use Turn on the power switch to see if the power lamp lights up Check if the battery has been charged 8 Power cord and plug z z Before use Check the pow...

Page 43: ... time years Battery 2 Control unit 4 to 6 Caster 5 to 7 Brake rubber 3 to 5 Power cord 5 to 7 5 3 Version information of the software The version information for the control software which is installed in the operating table can be verified via the label that is directly applied on the on board microcomputer CAUTION The version information of the software is mainly for the service and maintenance ...

Page 44: ...e Each for left and right 90 Detachment Head plate Leg plate left right Others Emergency brake release dial Rating Classification as per Class1 device internal power source device Note 2 Rated voltage AC 100 240 V Frequency 50 60 Hz Battery power DC 24 V Power input 400 VA Operating voltage DC 5 V 24 V Duty cycle 3 min on 7 min off Note 3 Others Recovery from defibrillator is within five seconds C...

Page 45: ...r s line are aligned Note 2 When the battery power is used Note 3 Consecutive pressing of the switch of the control unit Note 4 Excluding the side rail Note 5 Rough dimension Note 6 Company standard in case that appropriate maintenance and inspection is done Note 7 Based on Mizuho s own validation data Note 8 IEC 60601 1 2012 Medical electrical equipment Part1 General requirements for safety 6 2 E...

Page 46: ...ly stop the operation Otherwise the operating table may get damaged z z Use the extra switch to operate the operating table 1 Open the lid of the extra switch Control panel 2 Press any function button on the control panel according to the desired direction Reverse Trendelenburg Head up Lateral tilt Right down Back plate bending Up Lift Up Trendelenburg Head down Lateral tilt Left down Back plate b...

Page 47: ...elease the brake 1 Turn the emergency brake release dial toward UNLOCK The brake is released Turn the dial WARNING To move the operating table after you released the brake activate the brake again with the following procedure If the operating table is moved with the brake released the patient may fall from the operating table 1 Return the emergency brake release dial to LOCK direction Turn the dia...

Page 48: ... function The emergency brake release dial is in the release UNLOCK position Turn the emergency brake release dial toward LOCK à Page 45 The brake has not been unlocked with the control unit since the brake was released with the emergency brake release dial Use the control unit to unlock the brake à Page 45 The back plate cannot be flexed below the level position The tabletop is slid in the foot d...

Page 49: ...sure to clean and disinfect the operating table when requesting to have it repaired Maintenance by providers For safety use of this product make sure to perform the periodical inspection by Mizuho or the certified provider once a year Inspections and maintenances by other than Mizuho or the certified provider could cause any adverse event such as deterioration of the performance and functions For ...

Page 50: ...served to verify normal operation in the configuration in which it will be used GUIDANCE AND MANUFACTURER S DECLARATION ELECTROMAGNETIC COMPATIBILITY Operating Table MOT 5602BW is intended for use in the electromagnetic environment specified below The customer or the user of Operating Table MOT 5602BW should assure that it is used in such an environment Emissions test Compliance Electromagnetic en...

Page 51: ...ximum output power of transmitter Separation distance according to frequency of transmitter m 150 kHz to 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz W d 1 2 P d 1 2 P d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power not listed above the recommended separation distance d in metres m can be estimated using the e...

Page 52: ...line s to line s 2 kV line s to earth 1 kV line s to line s 2 kV line s to earth Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec 5 UT 95 dip ...

Page 53: ...m Field strengths from fixed RF transmitters as determined by an electromagnetic site survey a should be less than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80 MHz and 800 MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is af...

Page 54: ...p or the head down position and the entire tabletop moves to the Center up or Center down Lateral tilt Tabletop of the operating table moves to the left down or the right down position in the view from the head side Tabletop The light blue portion of the figure below Trendelenburg Tabletop of the operating table moves to the head up or the head down position ...

Page 55: ...Revision Record 10 04 2017 REV1 New release 05 2019 Version 2 Revision ...

Page 56: ...Sale Agent Autorized Representative Europe Manufacturer 3 30 13 Hongo Bunkyo ku Tokyo 113 0033 Japan Emergo Europe B V Prinsessegracht 20 2514 AP The Hague The Netherlands http www mizuho co jp ...

Reviews: