SECTION 2: OVERVIEW
21
3.4 Feed Water
(*)
Tube and Drain Tube Connection
Feed water
(*)
tube connection diameter Connecting
φ
6 mm OD tube
(*10)
Drain tube connection diameter Connecting
φ
8 mm OD tube
(*10)
*1: When using the device in North America, be sure to use the provided AC adapter.
Manufacturer: GlobTek, Inc.
Model: GTM96300-3024-R3A
*2: The AC adapter is not provided with the device sold outside North America.
Prepare and use an AC adapter that is compliant with the laws and regulations of the country
where the device is used and that satisfies the following conditions:
Output voltage: 24 VDC
Output current: 1 A or more
Plug shape:
φ
5.5 mm OD,
φ
2.5 mm ID
Plug polarity: Center positive (shown in the figure
to the right)
*3: The wall-mounting bracket is provided with the device.
*4: Indicated value for evaluation results:
0 (less than 1 mg/L), 1 (1 mg/L or more but less than 2 mg/L),
2 (2 mg/L or more but less than 3 mg/L), 3 (3 mg/L or more but less than 5 mg/L),
>5 (5 mg/L or more)
*5: The warranty period of a reagent cartridge is 1 year for unopened product or 4 months after
opening the pack. The reagent will last approximately 4 months on the basis of an hourly
monitoring. If monitoring is taken every 30 minutes or repeated frequently, then the reagent
may not last for 4 months.
*6: The LEDs on the panel turn on when a remote alarm or concentration alarm occurs or when
the reagent cartridge needs replacement. Once the device alarm and hardness
(*)
leakage are
resolved, or the reagent cartridge is replaced, each LED turn off.
*7: The concentration alarm and device alarm are output together.
*8: The concentration alarm is output independently.
*9: The drain water flow rate when with a constant flow valve is installed is shown in the table.
The drain water flow rate may vary depending on the water temperature or degradation level
of a constant flow valve.
*10: The feed water
(*)
tube and drain tube are provided with the device.
Replace a tube if a defect such as deformation (a bend or fold), discoloration, hardening, or
cracking is detected on the tube surface. If a deteriorated tube is continuously used, it can rip
and cause water leakage. Make a daily check on whether the tubes are in normal condition.
Use specified tubes when replacing tubes.
Using tubes other than specified tubes may cause leakage. Contact your dealer or MIURA
sales office.