background image

2

mvip@mitech-ndt.com

1.Overview

MITECH portable vibration meter MV800C is made up of piezoelectric transducer and digital display

circuit. It is mainly used to measure vibration acceleration, speed and displacement of various rotary

machine. This product has the advantages of large measurement, wide variety, intuitively demonstration,

small size, light weight etc. It is widely used in measurement on site or at height and can reduce the

working strength, improve the working efficiency.

1.1 Application

MV800 is mainly used in monitoring the vibration of machines, such as industries of manufacturing,

electric, metallurgy, vehicle etc, specific on the quick check and preventive check of bearing, gear, motor,

fan lathe etc.

1.2 Features

Separate design, small size, stable performance low-consumption,attractive appearance.

Liquid crystal display, humanized interface, easy to operate.

Apply to harsh environment, with anti-electromagnetic interference ability, high accuracy.

High frequency and low frequency measurement, easy to judge fault type of machine.

Equip reliable ring accelerometer, superior than compression type sensor

Auto sleep and auto power off

In accordance with national standard, the measurement value can be compared with ISO2372 to

judge the running state of device.

Signal output, equips with earphone can realize the function of stethoscope.

Signal output equips with oscillometer to monitor and record the vibration.

1.3 Technical Parameters

Vibration pickup: Integrated ring acceleration transducer

Measurement range(displacement): 0.001~1.999mm(Peak to peak: valid value*2√2)

Measurement range(speed): 0.1~19939mm/s (valid value)

Measurement range(acceleration): 0.1~199.9m/s2(peak value: valid value*√2)

Measurement accuracy:

±

5% measured value

Indication error:

±

2

Frequency range: 10Hz~10KHz

Display mode: 3-digit Semi liquid crystal display

Sampling period: 1s

Output signal: AC 2v(peak value, full measurement above 10k)

Dimensions: 185mm*68mm*30mm

Power model: 6F22 9v packed cells

Battery life: continuous use for 30 hours

1.4 Working Principle

MV800C combines the quartz crystal and PZT piezoelectric effect together. When the quartz crystal or

PZT is under strength, the surface generates charger, the vibration was changed into electrical signal by

acceleration transducer. After processing and analyzing, the acceleration, speed and displacement are

shown out in the liquid screen.

1.5 Working Conditions

Operation temperature: 0~50

Memory temperature: -30

~+70

Relative humidity: below 90%RH

[email protected]

ponpe.com

Summary of Contents for MV800C

Page 1: ...MITECH CO LTD www mitech ndt com www ponpe com ...

Page 2: ... 2 Structure and Outline 3 2 1 Main Unit 3 2 2 Display 3 3 Keypad Definition 3 4 Operation 3 4 1 Preparations Before Using 3 4 2 Selection of Probe 4 4 3 Frequency Switch 4 4 4 Set Vibration Parameters 4 4 5 Set Frequency Range 5 4 6 Signal Output 5 4 7 Troubleshooting 5 5 Cautions 5 6 The Standard of Vibration Intensity 5 7 Maintenance Notes 6 User Notes 7 sale ponpe com ...

Page 3: ...ccordance with national standard the measurement value can be compared with ISO2372 to judge the running state of device Signal output equips with earphone can realize the function of stethoscope Signal output equips with oscillometer to monitor and record the vibration 1 3 Technical Parameters Vibration pickup Integrated ring acceleration transducer Measurement range displacement 0 001 1 999mm Pe...

Page 4: ...do the new test 4 Release the measurement key for about 1 minute it will auto power off 4 Operation 4 1 Preparations Before Using 1 Install battery Open the cover put in the battery 1 Check the battery voltage Configuration NO Part name QTY Remarks Standard config 1 Main unit 1 2 Short probe 1 3 6F22 battery 1 4 User s manual 1 5 Warranty 1 Optional config 6 Long probe 1 7 Earphone 1 8 Sheathe 1 9...

Page 5: ...there is a flat frequency above 1KHz there is no need to use probe Short probe Short probe is necessary for frequency of 10Hz 10KHz It is more reliable Long probe When the short probe is hard to test use 8cm long probe In low frequency of 10 1KHz long probe has good response above 1KHz the response will reduce No probe It is ok when the surface is well connected if the sensor and workpiece is not ...

Page 6: ...means the vibration is caused be high frequency such as the deterioration of rolling bearing gear tooth broken etc 5 Cautions When using integrated vibration meter the strength on probe should between 5N 10N Pay attention to the right angle of probe When using different probes the results might be different especially in high frequency Do not use pincers to avoid damage Theoretically MV800C can te...

Page 7: ...is almost out please change the battery Take out the battery when not using to avoid the battery leakage After using clean the tester with soft cloth avoid rain or oil infiltration When there happens abnormal things please do not self dismantle the tester deliver it to our company sale ponpe com ponpe com ...

Page 8: ...eporting to repair sheet In accordance with the international relevant regulations the following are not within the scope of free warranty Damage caused by man made or improper keeping Self dismantle or non special repair shop dismantle Do not follow the requirement of service registration or warranty expired Consumable parts Service promise MITECH users have lifelong maintenance service Free main...