Chapter 5 Setting Up the Unit
Exiting the Source Finder Mode
To exit the Source Finder Mode:
•
Press and hold the
RATE
key until the
display changes to a flashing
0
.
•
Release the
RATE
key.
The Sleep Mode
When the Sleep Mode is enabled, most of the
unit’s circuits will be turned off to conserve
battery power. Every 5 minutes, the unit will
wake up for 4–6 seconds and check the
current Rate.
The battery life time for the UltraRadiac-Plus
in this mode is approximately 2000 hours.
WARNING:
Do not place the unit in
Sleep Mode before
responding!
Doing so can expose you to
radiation for as long as five
minutes before you are
alerted.