Chapter 5 Setting Up the Unit
•
Press the
CLR/TEST
key; the flashing
display will show the selected (
H
or
L
)
alarm threshold setting.
•
If the optional Vibrator Alarm has been
activated, the display will alternate
between the threshold value and a
vertically scrolling horizontal bar
(–
–
–).
Changing the Dose Alarm Indicators
To change the High level (H) or Low level (L)
alarm indicator settings:
•
Press the
ALARM
and
DOSE
keys.
•
You’ll see an
H
, for High Level Alarm,
displayed.
•
To change the Low Level indicators,
press the
DOSE
key again; the display
will show an
L
.
•
Press the
CLR/TEST
key to access the
selected (
H
or
L
) indicators.
•
Repeatedly press the
ALARM
key until
you see the desired combination of
indicators, AUDio, VISual, vibrator
(–
–
–), or no indicators.