![MIR Spirolab User Manual Download Page 31](http://html1.mh-extra.com/html/mir/spirolab/spirolab_user-manual_1804906031.webp)
30
Technical Specification
Method of detection
Red and infra-red light absorption
Measuring range of %SpO2
0 – 99% (with 1% increments)
Resolution of SpO2
1%
Accuracy of %SpO2
2% between 70-99% SpO2
Number of beats for calculating the
median SpO2 %
8 beats
Pulse rate measuring range
18 – 300 BPM
Resolution of pulse rate
1 BPM
Pulse rate accuracy
2 BPM or 2% of highest value
Interval for calculating median pulse
rate
8 seconds
Signal quality
0 - 8 display segments
Oximetry definitions
Oximetry specifications
Desaturation event
Drop in SpO2 >= 4% in a 8- 40 sec period limited and subsequent rise >=2% inside an overall period of 150 sec
Pulse variation event
Rise in Pulse >= 10 BPM in a 8- 40 sec limited period and subsequent drop >=8 BPM inside an overall period of 150 sec.