Optosense LLC
USER MANUAL
Design department
MIPEX-04 ULTRA-LOW POWER INFRARED GAS SENSOR
ESAT.100000.00.02 UM
Revision 02
December 12th, 2016
Page 29 of 41
THE SOLE PROPERTY OF OPTOSENSE LLC. ANY REPRODUCTION WITHOUT THE WRITTEN PERMISSION OF OPTOSENSE LLC IS PROHIBITED.
Table 11. Bits of status
bytes’ description
Bit
number
Status word
compliance
9
Description
Recommendations
–
00
Normal mode, static temperature
–
0
10
Sensor is warming up.
Do not perform span calibration.
1
50
An abrupt signal change (due to feed-
ing a gas mixture to sensor atmos-
phere) or increased noise on sensor
optical elements.
Do not perform span calibration or zeroing. If
this status longer than for 20 minutes at sta-
ble conditions, replace sensor.
2
30
One of the signal values is lower than it
is allowed.
Possible exposure to moisture. Dry sensor.
After drying, if this status lasts longer than
for 20 minutes at stable conditions, replace
sensor.
3
–
reserved
Does not contain any significant information.
4
21
Dynamic temperature mode (tempera-
ture changes faster than by 0.6 °C/min).
Do not perform span calibration or zeroing.
5
22
Dynamic temperature mode (tempera-
ture changes faster than by 2 °C/min).
Do not perform span calibration or zeroing.
6
40
Exceeding temperature limits.
Check the ambient temperature.
7
90
Firmware failure (flash memory issues).
Contact support for details.
8
11
Data request rate is more than 1 Hz.
Do not perform span calibration or zeroing.
Decrease request rate.
9
31
Ratio S
tz0
(see Table 15) exceeds
allowable upper limit.
Possible exposure to moisture. After drying,
if this status lasts longer than for 20 minutes
at stable conditions, perform zeroing (see
Appendix D.1 for procedure description).
10
–
reserved
Does not contain any significant information.
11
51
Complex status. Technological failure.
Contact support for details.
12
–
reserved
Does not contain any significant information.
13
–
reserved
Does not contain any significant information.
14
–
reserved
Does not contain any significant information.
15
–
reserved
Does not contain any significant information.
(4
∨
5) & 9
24
Dynamic temperature mode; zero shifts
to negative value.
Do not perform span calibration or zeroing.
9
Status word is contained in response to command <F> (see Table 15).