Table 15. Structure of response to <F> command
Command <F> (hex: “46 0D”) response
Byte number
Byte data
Data description
1
Hex: “0x0E”
Special character
2-6
T
Sensor temperature expressed in ADC counts
7
Hex: “0x09”
Tabulation symbol
8-12
S
t
The ratio
U
s
U
ref
considering temperature correction
13
Hex: “0x09”
Tabulation symbol
14-18
U
s
Operating signal in ADC counts
19
Hex: “0x09”
Tabulation symbol
20-24
U
ref
Reference signal in ADC counts
25
Hex: “0x09”
Tabulation symbol
26-30
S
tz0
The ratio S
t
considering ZERO and ZERO2 coefficients
31
Hex: “0x09”
Tabulation symbol
32-36
S
tz
The ratio S
tz0
considering the drift compensation algorithm
37
Hex: “0x09”
Tabulation symbol
38-42
S
tzkt
The ratio S
tz
considering the coefficient of temperature
sensitivity
43
Hex: “0x09”
Tabulation symbol
44-48
C
Concentration based on factory settings
49
Hex: “0x09”
Tabulation symbol
50-54
C
1
Concentration based on user settings
55
Hex: “0x09”
Tabulation symbol
56-60
Status word
Status word (see Table 16 for details)
61
Hex: “0x09”
Tabulation symbol
62-69
S/N
Sensor serial number
70
Hex: “0x09”
Tabulation symbol
71
CRC
CRC calculated using “exclusive OR” method
72
Hex: “0x09”
Tabulation symbol
73
Hex: “0x0D”
Carriage return