Optosense LLC
USER MANUAL
Design department
SMALL-SIZE EXPLOSIVE GAS MEASURING
SENSOR MIPEX-02-X-X-X.1 X (RX)
ESAT.413347.005 UM
Revision 3.0
January 19th, 2017
Page 36 of 46
THE SOLE PROPERTY OF OPTOSENSE LLC. ANY REPRODUCTION WITHOUT THE WRITTEN PERMISSION OF OPTOSENSE LLC IS PROHIBITED.
Request syntax
USER
mode
OEM
mode
Response description
Command description
DATEZC?
yes
yes
DD.MM.YY
Returns date of the last span
calibration operation.
DATEZC DD.MM.YY
no
yes
DATEZC DD.MM.YY OK
or
DATEZC DD.MM.YY FAULT
Writes the last date of span
calibration to sensor memory
(DD
– 00…31; MM – 0…12;
YY
– 00…99).
INIT
no
yes
INIT OK
or
INIT FAULT
Resets sensor to the factory
calibration settings. It initializes
zero and scale coefficients
setting them to one (“1”). After
sending this command, it is
recommended to check sensor
readings. In case if sensor
does
not
comply
to
its
metrological
specifications,
perform span calibration (see
Appendix D.1)
SETC XXXXX
yes
yes
SETC XXXXX OK
or
SETC XXXXX FAULT
Intended for determining the
current ambient temperature in
Celsius
degrees
(XXXXX
stands for degrees; i.e. 00023
= 23 °C). Sensor writes current
temperature ADC value to its
memory.
UPLOAD
no
yes
U_OK
By this command, sensor is
set to firmware updating mode.
This mode switches off auto-
matically upon the firmware is
updated. To switch it off manu-
ally reset sensor power supply.
ZERO
no
yes
ZERO OK
or
ZERO FAULT
Writes
current
temperature
ADC value and ratio S
t
value
to sensor controller memory by
this command.
ZERO0
no
yes
ZERO0 OK
or
ZERO0 FAULT
Erases
temperature
coeffi-
cients previously defined by
user.
ZERO2
no
yes
ZERO2 OK
or
ZERO2 FAULT
Intended for zeroing sensor. It
initiates calculation and storing
of the unifying zero coefficient
within sensor memory. C and
C
1
values are zeroed and ratio
S
tz
becomes equal 1 by this
command.