SMALL-SIZE EXPLOSIVE GAS MEASURING SENSOR MIPEX-02-
Х-X-X.1 X
File name: ESAT.413347.006 UM v.2.0.docx
Revision 2.0 27 August 2015
Page. 31 of 38
THE SOLE PROPERTY OF OPTOSENSE LLC.
ANY REPRODUCTION
WITHOUT THE WRITTEN PERMISSION OF OPTOSENSE LLC IS PROHIBITED.
SETC
XXXXX
No
One-point temperature calibration.
SETC XXXXX
OK
XXXXX
– temperature in Celsius degrees. With
SETC
command,
sensor stores in memory the ambient temperature value. Also see
CCS
,
CFS
,
CKS
command description.
ZERO
No
Setting zero in current temperature.
ZERO OK
Command allows to set zero in different points of operating
temperature range. See detailed procedure below.
ZERO
XXXXX
YYYYY
No
Manual setting of temperature
coefficients.
ZERO
XXXXX
YYYYY OK
XXXXX
– temperature data,
YYYYY
– signal ratio
<Stz>
.
ZERO2
No
Setting zero in whole operating
temperature range.
ZERO2 OK
The command initiate calculation and store a unifying zero coefficient
to the memory. In case of supplying this command, concentration
Conc
= 0,
Conc1
= 0, relation
Stz
is corrected to 10000 (see Table
INIT
No
Restore to factory default.
INIT OK
or
INIT FAULT
The command is used in case of wrong sensor calibration or other
improper actions over the sensor.