Table 7. Structure of F command response
Byte number
Byte data
Data description
1
0Eh
Special character
2-6
Termo
Sensor temperature in ADC readings
7
09h
Tabulation
8-12
<St>
Ratio Us/Uref with temperature index
13
09h
Tabulation
14-18
Us
Operating signal in ADC readings
19
09h
Tabulation
20-24
Uref
Reference signal in ADC readings
25
09h
Tabulation
26-30
Stz0
Ratio <Stz> with manual zeroing settings (ZERO)
31
09h
Tabulation
32-36
Stz
Ratio Stz0 with drift compensation algorithm
37
09h
Tabulation
38-42
Stzkt
Ratio Stz with coefficient of temperature sensitivity
43
09h
Tabulation
44-48
Conc
Concentration (according to a factory calibration)
49
09h
Tabulation
50-54
Conc1
Concentration scaled by user
55
09h
Tabulation
56-60
Status
Status word (see Table 9)
61
09h
Tabulation
62-71
Sn
Sensor serial number
70
09h
Tabulation
71
ControlSum
Checksum based on XOR
72
09h
Tabulation
73
0Dh
Carriage return
Table 8. Structure of @ and @*X commands
Command @ (40h 0Dh)
response, byte number and data
Command @*x (40 2A x 0D) response, byte number
and data
1
2
1
2
3
Conc1H
Conc1L
@(40h)
Conc1H
Conc1L