4.3.Calibration
31
4
Calibration
&RQ¿UPWKHVHWWLQJV
&RQ¿UPWKHYDOXHRI>:(,*+,1*&$3$&,7<@
The setting is completed if the value set by the step
2) appears under [WEIGHING CAPACITY].
POINT
- To stabilize the display of weighing values, enter the settings so that the display
resolution ([WEIGHING CAPACITY] ÷ [DIVISION]) becomes 10 000 or less.
4.3.4. Zero adjustment
Register the load cell output value when the load is zero.
The following two methods are available for [ZERO ADJUSTMENT].
- By reading the weighing value (initial load value)
Read the load cell output value with no load on the weighing device to register it as the
zero point.
- By entering the numerical value of the output voltage
Enter the load cell output voltage value when the load is zero to register it as the zero
point.
In normal operation environments, perform [ZERO ADJUSTMENT] by reading the weighing
value (initial load value).
Perform [ZERO ADJUSTMENT] by entering the numerical value of the output voltage value
only when an object is on the weighing device and you cannot remove it (you cannot prepare
an initial sate including the weight of an empty container).
(1) Read the weighing value (initial load value)
Before [ZERO ADJUSTMENT], make sure that no load is placed on the weighing device.
1) Perform [ZERO ADJUSTMENT]
1. Tap [ZERO ADJUSTMENT] on [CALIBRATION 1/2].
The screen for [ZERO ADJUSTMENT] will be
displayed.
2. Tap [OK].
The message [EXECUTE ZERO CALIBRATION?]
will be displayed.