2.2. Wiring of strain gage based transducers
8
W
iring
2
2.2.2. 4-wires type wiring cable
Connect CSD-912B with a load cell using a 4-wires cable as follows.
POINT
- Short out A-F and C-G of the strain gage based transducer terminal board using
the attached short bar. Without shorting out the F and G terminal, CSD-912B
does not operate normally.
- If the 4-wires cable is 30 m or more, the resistance of the cable causes decline in
the input voltage of CSD-912B, and the accuracy may not be covered under
warranty.
/RDGFHOOH[FLWDWLRQ5('
/RDGFHOOH[FL7
/RDGFHOOVLJQDO*51
/RDGFHOOVLJQDO%/8
6KLHOG<(/
Attached short bar
/RDGFHOO
(;&
SIG-
SIG+
(;&